      "status": "Terminated",
      "city": "Salt Lake City",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82354",
      "recalling_firm": "BioFire Diagnostics, LLC",
      "address_1": "515 S Colorow Dr",
      "address_2": "N/A",
      "postal_code": "84108-1248",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "MA, NH, VT, CT, NJ, NY, PA, DE, DC, VA, MD, WV, NC, SC, GA, FL, AL, TN, MS, KY, OH, IN, MI, IA, WI, MN, SD, ND, MT, IL, MO, KS, NE, LA, AR, OK, TX, CO, WY, ID, UT, AZ, NM, NV, CA, HI, OR, WA, AK, BM",
      "recall_number": "Z-1133-2019",
      "product_description": "FilmArray BCID Panel, IVD, Rx Only, BioFire Diagnostics,  REF: RFIT-ASY-0126, and RFIT-ASY-0127,  when used with the following blood culture bottles: Bottle Description/Part Number: BACT/ALERT FA Plus/410851; BACT/ALERT FN Plus/410852; BACT/ALERT PF Plus/410853",
      "product_quantity": "20350",
      "reason_for_recall": "Increased risk of false positive Enterobacteriaceae and Escherichia coli results on affected blood culture identification panels when using certain blood culture bottles.",
      "recall_initiation_date": "20190307",
      "center_classification_date": "20190412",
      "termination_date": "20220824",
      "report_date": "20190424",
      "code_info": "All lots of BCID Panels, when used with the following blood culture bottles, Part Number/Lot Numbers:410851/4052663, 4052639, 4052640, 4052599, 4052598, 4052547, 4052546, 4052471, 4052472, 4052799, 4052800, 4052815, 4052816; 410852/4052871, 4052872, 4052459, 4052458, 4052565, 4052564, 4052705, 4052704, 4052735, 4052617, 405261; 410853/4052717, 4052718, 4052503, 4052504"
    }
  ]
}