  "results": [
    {
      "status": "Terminated",
      "city": "Lawrence",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79276",
      "recalling_firm": "NxStage Medical, Inc.",
      "address_1": "350 Merrimack Street",
      "address_2": "N/A",
      "postal_code": "01843-1748",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US nationwide distribution",
      "recall_number": "Z-1132-2018",
      "product_description": "NxStage PureFlow B Solution - Premixed Dialysate for Hemodialysis with NxStage System One",
      "product_quantity": "1243400 units",
      "reason_for_recall": "Using different fluid formulations of NxStage PureFlow B Solution at the same time during treatment may cause patient health risk.  The NxStage PureFlow B Solution is for use with systems that use one premixed dialysate during hemodialysis. The original labeling did not warn against mixing dialysate formulations and did not instruct staff to verify that the fluid is flowing during use.",
      "recall_initiation_date": "20171110",
      "center_classification_date": "20180319",
      "termination_date": "20191206",
      "report_date": "20180328",
      "code_info": "RFP-400 -  M535RFP4000  RFP-400-BB - 00842289101258  RFP-401 M - 535RPF4010  RFP-401-BB  - 00842289101265  RFP-402 - M535RFP4020  RFP-403 - M535RFP4030  RFP-404-  M535RFP4040  RFP-405 - M535RFP4050  RFP-406 - M535RFP4060  RFP-453 - M535RFP4530  RFP-454 - M535RFP4540  RFP-455 - M535RFP4550  RFP-456 - M535RFP4560"
    }
  ]
}