  "results": [
    {
      "status": "Terminated",
      "city": "Miami Lakes",
      "state": "FL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70340",
      "recalling_firm": "HeartWare Inc",
      "address_1": "14400 Nw 60th Ave",
      "address_2": "N/A",
      "postal_code": "33014-2807",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of:  AL, AZ, CA, CO, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NY, OH, OR, PA, TX, UT, VA, WA and District of Columbia and countries of:  Canada, Poland, Australia, Austria, Italy, DE, New Zealand, South Africa, France, Switzerland, Germany, Brazil, UK, Belgium, Netherlands, Israel, Malaysia, Turkey, Greece, Japan, Singapore, Poland, Lithuania, Sweden, Norway, Finland, and India.",
      "recall_number": "Z-1131-2015",
      "product_description": "HeartWare Ventricular Assist System (HeartWare Controller)    Product Usage:  For use as a bridge-to-cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure. The HeartWare System is designed for in-hospital and out-of-hospital settings, including transportation via fix-wing aircraft or helicopter.",
      "product_quantity": "4,845",
      "reason_for_recall": "The affected clinical trial Controllers exhibit a higher susceptibility to ESD than newer commercial Controllers.",
      "recall_initiation_date": "20150116",
      "center_classification_date": "20150225",
      "termination_date": "20200205",
      "report_date": "20150304",
      "code_info": "Product Codes- US product: Model No.: 1400, 1401US; International product: Model No.: 1400, 1401XX ('XX' represents country designation)  Serial No.: CON000001 through CON005472",
      "more_code_info": ""
    }
  ]
}