{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Milpitas",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64478",
      "recalling_firm": "Lifescan Inc",
      "address_1": "1000 Gibraltar Dr",
      "address_2": "N/A",
      "postal_code": "95035-6301",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, 10, IL, IN, KS, KY, LA, MA, MD, ME, Ml, MN, MO, MS, MT, NC, NO, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SO, TN, TX, UT, VA, VI, VT, WA, WI,WV, and WY, and the countries of Russia and Ukraine.",
      "recall_number": "Z-1129-2013",
      "product_description": "LifeScan brand OneTouch¿ Ultra Control Solution, Part Numbers/Model Numbers: 01045808, 02141603.    Testing of Blood Glucose Levels.",
      "product_quantity": "132,301 units total",
      "reason_for_recall": "When OneTouch¿ Ultra and OneTouch¿ Select test strips are tested with Ultra and Select control solution the meter may display a result outside the control solution test range printed on the strip vial label even though the system is working properly.",
      "recall_initiation_date": "20130225",
      "center_classification_date": "20130417",
      "termination_date": "20131023",
      "report_date": "20130424",
      "code_info": "2AA2G01, Exp. Date: 7-31-2014;  2AA2G02, Exp. Date: 7-31-2014;  2AA2G03, Exp. Date: 8-31-2014;  3AA2G04, Exp. Date: 12-31-2014;  3AA2G05, Exp. Date: 12-31-2014"
    }
  ]
}