  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Plantation",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70374",
      "recalling_firm": "Mako Surgical Corporation",
      "address_1": "2555 Davie Rd Ste 110",
      "address_2": "",
      "postal_code": "33317-7424",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution-including the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, LA, MA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV and Hawaii.",
      "recall_number": "Z-1128-2015",
      "product_description": "Restoris Partial Knee Application (PKA) RIO (TGS 2.0).    For use with the Robotic Arm Interactive Orthopedic System (RIO), is indicated for use in surgical knee procedures.",
      "product_quantity": "228 units",
      "reason_for_recall": "When using the MAKOplasty partial knee Arthroplasty application, the burr continues spinning outside of the stereotactic boundary and after the control switches (foot pedal and trigger) cease to be activated",
      "recall_initiation_date": "20150122",
      "center_classification_date": "20150219",
      "termination_date": "20161214",
      "report_date": "20150225",
      "code_info": "All lots, Part Numbers: 201000, 203999, 207300, and 209930.",
      "more_code_info": ""
    }
  ]
}