  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Tonawanda",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64698",
      "recalling_firm": "Sr Instruments Inc",
      "address_1": "600 Young St",
      "address_2": "N/A",
      "postal_code": "14150",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "USA Nationwide Distribution in the state of Texas.",
      "recall_number": "Z-1128-2013",
      "product_description": "SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters.  Model# FG3105 Patient Scales.      Product Usage:  The FG3105 Scale is designed for Joerns Healthcare for use in portable patient weighing systems. The scale is designed to be connected between the patient hoist mechanism and patient sling apparatus. The scale is not intended to be used for patient transport.",
      "product_quantity": "806 patient scales",
      "reason_for_recall": "SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they received two incident reports regarding two field failures of a tensile type load cell involving the FG3105 Patient Scale.  The failures resulted in patient falls.",
      "recall_initiation_date": "20121213",
      "center_classification_date": "20130416",
      "termination_date": "20150312",
      "report_date": "20130424",
      "code_info": "Serial numbers: 1005 thru 1780, 1785 thru 1809, 1866 thru 1868 and 1870 thru 1875."
    }
  ]
}