  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Mercedita",
      "address_1": "Turpeaux Industrial Park Rd 1 km 123.4",
      "reason_for_recall": "As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013.  The affected products are sterile.",
      "address_2": "",
      "product_quantity": "0",
      "code_info": "lot no.:  62051735  62221046  62262799  62272099  62279770  62279770N  62079181  62162466  62185965  62185966  62221047  62262801  62262802  62262803  62279771  62285361  62311660  62393063  62279771N  62285361N  62311660N  61520812  62162464  62221048  62279772  62393064  62279772N  62393064V",
      "center_classification_date": "20160312",
      "distribution_pattern": "US Nationwide Distribution",
      "state": "PR",
      "product_description": "Product 29 consists of all product under  product code: LPHand same usage:  Item no:\t  32902603800\tFEMORAL HEAD +0.5 X 22 MM  32902604810\tFEMORAL HEAD +3.5 X 22 MM  32902604845\tFEMORAL HEAD +7.0 X 22 MM    For use in total or hemi hip arthroplasty",
      "report_date": "20160323",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Manufacturing B.V.",
      "recall_number": "Z-1127-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "72900",
      "termination_date": "20170413",
      "more_code_info": "",
      "recall_initiation_date": "20160111",
      "postal_code": "00715",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}