      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64712",
      "recalling_firm": "Accuray Incorporated",
      "address_1": "1310 Chesapeake Ter",
      "address_2": "N/A",
      "postal_code": "94089-1100",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution, including Nationwide in the US and the countries of Korea, Myanmar, Canada, Belgium, England, Czech Republic, Finland, France, Germany, Greece, India, Italy, Poland, Russia, Spain, Switzerland, Turkey, Columbia, Mexico, Japan and Saudi Arabia.",
      "recall_number": "Z-1126-2013",
      "product_description": "Accuray CyberKnife System Medical Physicist, Robotic Radiosurgery System, Medical charged-particle radiation therapy system, Model: Iris Variable Aperture Collimator P/Ns 028986 and 031777. The CyberKnife System is indicated for radiation treatment.",
      "product_quantity": "176 units",
      "reason_for_recall": "A defect on the master-side connector block of the pneumatic tool changer could introduce a tilt in the collimator housings, resulting in a potential shift to the central axis of the radiation beam. The issue could pose significant clinical impact to plans created using the smallest collimators.",
      "recall_initiation_date": "20130319",
      "center_classification_date": "20130416",
      "termination_date": "20131023",
      "report_date": "20130424",
      "code_info": "All units of these models."
    }
  ]
}