      "country": "United States",
      "city": "Mercedita",
      "address_1": "Turpeaux Industrial Park Rd 1 km 123.4",
      "reason_for_recall": "As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013.  The affected products are sterile.",
      "address_2": "",
      "product_quantity": "0",
      "code_info": "lot no.:  61555543  61597965  61670825  62079145  62079148  62124066  62135039  62185960  62203621  62221041  62221043  62221044  62267947  62267949  62267950  62268526  62277492  62277493  62278701  62278702  62278703  62340123  62397828  62397831  62267950N  62278701N  62278702N  62278703N  61597959  61653581  62218170  62265345  62277494  62294359  62218170N  62277494N  62294359N  61292730N  62122127  62267951  62267952  62278704  62267952N  62278704N  62181482  62304835  62304835N  62203622  62203622N  62289044  62391711  62289044N  62391711V  61854615  61977887  62135046  62162460  62267946  62268527  62340121  62268527N  62289054  62289083  62304837  62289054N  62289083N  62304837N",
      "center_classification_date": "20160312",
      "distribution_pattern": "US Nationwide Distribution",
      "state": "PR",
      "product_description": "Product 26 consists of all product under  product code: JWH and same usage:  Item no:\t  234703104\tDISTAL MEDIAL FEMORAL CON  234703206\tDISTAL MEDIAL FEMORAL CON  234703208\tDISTAL MEDIAL FEMORAL CON  234800504\tDISTAL POSTERIOR/LATERAL  234800506\tDISTAL POSTERIOR/LATERAL  234800508\tDISTAL POSTERIOR/LATERAL  234800510\tDISTAL POSTERIOR/LATERAL  234800604\tDISTAL POSTERIOR/LATERAL  234800606\tDISTAL POSTERIOR/LATERAL  234800608\tDISTAL POSTERIOR/LATERAL  234800610\tDISTAL POSTERIOR/LATERAL  234800614\tDISTAL POSTERIOR/LATERAL      Product Usage:  For use in total or hemi hip arthroplasty",
      "report_date": "20160323",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Manufacturing B.V.",
      "recall_number": "Z-1124-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "72900",
      "termination_date": "20170413",
      "more_code_info": "",
      "recall_initiation_date": "20160111",
      "postal_code": "00715",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}