  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bridgend South Wales",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64679",
      "recalling_firm": "Biomet U.K., Ltd.",
      "address_1": "Waterton Industrial Estate",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide), Europe, Japan, Australia, Canada, Singapore, South Africa, and Costa Rica.",
      "recall_number": "Z-1124-2013",
      "product_description": "REF 32-422760   Oxford Uni Knee System,Toffee Hammer,  non-sterile,   found in REF 32-422763 Oxford Mlcroplasty Knee System  Tibial instrument Tray     small surgical hammer supplied with the Oxford Partial Knee Microplasty Instrument set, used for impacting implants.",
      "product_quantity": "430",
      "reason_for_recall": "Investigation of complaints found excessive use of the instrument could result in cracking on the weld  between the head and the handle, which could lead to part of the head becoming detached from the handle.",
      "recall_initiation_date": "20130320",
      "center_classification_date": "20130415",
      "termination_date": "20150504",
      "report_date": "20130424",
      "code_info": "32-422760 ( sold as part of instrument set 32-422763 )  ZB100901,ZB110201,ZB110301,ZB110302,ZB110303,  ZB110304,ZB110601,ZB110602,ZB110701,ZB111201,  ZB111202,ZB120201,ZB120202,ZB120701,ZB120801,  ZB120802,ZB121102,ZB121103"
    }
  ]
}