  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Franklin Lakes",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82317",
      "recalling_firm": "Becton Dickinson & Company",
      "address_1": "1 Becton Dr",
      "address_2": "N/A",
      "postal_code": "07417-1815",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US distribution to the states of FL and NJ.",
      "recall_number": "Z-1122-2019",
      "product_description": "BD Blood Collection Assembly with Male Luer Lock, Catalog Number MBC6010",
      "product_quantity": "6,509,600",
      "reason_for_recall": "The luer could break and cause the device to leak or break off and get stuck in the fistula needle port rendering the port inaccessible for dialysis.  As a result, the patient would need to be re-cannulated with a new fistula needle to obtain their dialysis treatment.",
      "recall_initiation_date": "20190307",
      "center_classification_date": "20190410",
      "termination_date": "20200512",
      "report_date": "20190417",
      "code_info": "Lot Numbers, UDI: 18041501,  (17)210315(10)18041501, (01)50885403235915.     18042501,  (17)210325(10)18042501, (01)50885403235915.     18050501,  (17)210405(10)18050501, (01)50885403235915.     18051501,  (17)210415(10)18051501, (01)50885403235915.     18052501,  (17)210425(10)18052501, (01)50885403235915.     18060501, (17)210505(10)18060501, (01)50885403235915.     18061501,  (17)210515(10)18061501, (01)50885403235915.     18062501, (17)210525(10)18062501, (01)50885403235915.     18070501, (17)210605(10)18070501, (01)50885403235915.     18072501, (17)210625(10)18072501, (01)50885403235915.     18080501, (17)210705(10)18080501, (01)50885403235915.     18081501, (17)210715(10)18081501, (01)50885403235915."
    }
  ]
}