  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Goleta",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76249",
      "recalling_firm": "Medtronic Neurosurgery",
      "address_1": "125 Cremona Dr",
      "address_2": "N/A",
      "postal_code": "93117-3083",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of :    Algeria  India  Poland  Armenia  Indonesia  Portugal  Australia  Iran  Austria  Iraq  Qatar  Bangladesh  Ireland  Republic of Montenegro  Barbados  Israel  Republic of Serbia  Belarus  Italy  Romania  Belgium  Jamaica  Russian Federation  Bosnia and Herzegovina  Japan  Saudi Arabia  Botswana  Jordan  Singapore  Bulgaria  Kazakhstan  Slovakia  Cambodia  Kenya  South Africa  Canada  Kuwait  South Korea  Canary Islands  Latvia  Spain  Cayman Islands  Lebanon  Sri Lanka  China  Libya  Sweden  Croatia  Lithuania  Switzerland  Cyprus  Macau  Syria",
      "recall_number": "Z-1121-2017",
      "product_description": "Medtronic Strata II / Strata NSC Valves (includes Burr Hole and Lumbar Peritoneal.  All Catalog Numbers.    Strata-type valves are used in the management of hydrocephalus. The Strata II/Strata NSC valve is a shunt component designed to allow for continuous flow of cerebrospinal fluid (CSF) from the ventricles of the brain to the peritoneal cavity or right atrium of the heart. mechanism.       Strata Valve Magnet Reverse Polarity (Excludes StrataMR).    Catalog No.     27739  27740  27786  27787  27788  27789  27794  27812  27814  27815  27816  27817  27818  27820  27821  27822  27823  27824  27825  27827  27828  27830  27831  27832  27848  27864  27867  27868  27888  27903  27922  27823  27924  23042  27925  27926  27927  27932  27933  42335  42355  42365  42836  42856  42866  44420  44421  44430  44465  46070  46075  46080  46085  46635  46655  46665  46836  46856  46866  46871  46876  46881  46886  92355  92365  92856  92866  96655  96665",
      "product_quantity": "315,997",
      "reason_for_recall": "Medtronic Neurosurgery initiated a recall after receiving five (5) Product Experiences between February 15, 2015 and August 23, 2016 reporting of discrepancies between the pressure/performance level (PL) at which the valve was set and the readings determined via an x-ray image.",
      "recall_initiation_date": "20170117",
      "center_classification_date": "20170202",
      "termination_date": "20170602",
      "report_date": "20170208",
      "code_info": "All Lot Numbers",
      "more_code_info": ""
    }
  ]
}