  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Mercedita",
      "address_1": "Turpeaux Industrial Park Rd 1 km 123.4",
      "reason_for_recall": "As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013.  The affected products are sterile.",
      "address_2": "",
      "product_quantity": "16",
      "code_info": "lot no.:  61907682  62038367  62106797  62161479  62240743  62268577  62344683  62030932  62106796  62148557  62206538  62240742  62303507  62352823",
      "center_classification_date": "20160312",
      "distribution_pattern": "US Nationwide Distribution",
      "state": "PR",
      "product_description": "Product 22 consists of all product under   product code: JDI and same usage:  Item no:\t  811400218\tCPT 12/14 COCR REVISION S  811400318\tCPT 12/14 COCR REVISION S  811400420\tCPT 12/14 COCR REVISION S  811400423\tCPT 12/14 COCR REVISION S  811400426\tCPT 12/14 COCR REVISION S  811400010\tCPT 12/14 COCR SIZE 0 EXT  811400000\tCPT 12/14 COCR SIZE 0 STD  811400110\tCPT 12/14 COCR SIZE 1 EXT  811400100\tCPT 12/14 COCR SIZE 1 STD  811400210\tCPT 12/14 COCR SIZE 2 EXT  811400200\tCPT 12/14 COCR SIZE 2 STD  811400230\tCPT 12/14 COCR SIZE 2 XEX  811400310\tCPT 12/14 COCR SIZE 3 EXT  811400300\tCPT 12/14 COCR SIZE 3 STD  811400330\tCPT 12/14 COCR SIZE 3 XEX  811400410\tCPT 12/14 COCR SIZE 4 EXT  811400400\tCPT 12/14 COCR SIZE 4 STD  811400430\tCPT 12/14 COCR SIZE 4 XEX  811400510\tCPT 12/14 COCR SIZE 5 EXT  811400500\tCPT 12/14 COCR SIZE 5 STD  811400530\tCPT 12/14 COCR SIZE 5 XEX    For use in total or hemi hip arthroplasty",
      "report_date": "20160323",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Manufacturing B.V.",
      "recall_number": "Z-1120-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "72900",
      "termination_date": "20170413",
      "more_code_info": "",
      "recall_initiation_date": "20160111",
      "postal_code": "00715",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}