{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91483",
      "recalling_firm": "MEDLINE INDUSTRIES, LP - Northfield",
      "address_1": "3 Lakes Dr",
      "address_2": "N/A",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-1117-2023",
      "product_description": "Surgical convenience kits labeled as:  CENTURION MEDICAL PRODUCTS L & D TRAY, Reorder Number MNS9065",
      "product_quantity": "138 units",
      "reason_for_recall": "The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.",
      "recall_initiation_date": "20221207",
      "center_classification_date": "20230210",
      "report_date": "20230222",
      "code_info": "GTIN 40653160274385, Lot # 22HMF916"
    }
  ]
}