      "status": "Terminated",
      "city": "Crawley",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82384",
      "recalling_firm": "Elekta Limited",
      "address_1": "Linac House",
      "address_2": "Fleming Way",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of TX, NY, WI, and Denmark, Germany, Italy, Netherlands, Sweden, United Kingdom",
      "recall_number": "Z-1109-2019",
      "product_description": "Elekta Unity, Image-Guided Radiation Therapy System    Product Usage:  Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan. Elekta Unity is intended for use with compatible Treatment Planning and Oncology Information Systems.",
      "product_quantity": "12 units",
      "reason_for_recall": "Users need to be aware when using these protocols for daily on line plan adaptation that:  1) The images acquired using these protocols do not represent the average position of the anatomy during the respiratory motion cycle. The images are based on data acquired around full expiration.  2) The display of the images in the Elekta Unity Application software does not provide information about the protocol used to acquire the image eg. with or without respiratory triggering.  Users must select an appropriate scan protocol that is representative of the respiratory phase used in the reference plan.",
      "recall_initiation_date": "20190312",
      "center_classification_date": "20190408",
      "termination_date": "20231103",
      "report_date": "20190417",
      "code_info": "Unity Systems:  13553-006/600003, 13371-002/600021, 10817-011/600010, 10562-011/600016, 10719-T03001/600013, 30004070-001/600023, 10420-AVL-U/600008, 11014-UMCU-U/600007, 10157-006/600014, 11611-40/600011, 30001461-J001/600002, 11489-21/600009,"
    }
  ]
}