  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Turku",
      "state": "N/A",
      "country": "Finland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76165",
      "recalling_firm": "PerkinElmer Life and Analytical Sciences, Wallac, OY",
      "address_1": "Mustionkatu 6",
      "address_2": "P.O. Box 10",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "MA, SC, TX, AL  OUS: China, Mexico, Peru, Morocco, Italy, Finland, Ecuador, Brazil",
      "recall_number": "Z-1108-2017",
      "product_description": "Spare Disk Remover Module; Product Number: 61001820    Intended for in vitro quantitative and qualitative determination of analytes in body fluids.",
      "product_quantity": "24 instruments and 3 spare disk remover modules",
      "reason_for_recall": "The firm became aware that the ball bushing axels are defective in the GSP instrument. During exposure to humid conditions, there is a risk that the axle material will corrode. Deterioration of the axle can be observed as deposits of corrosion on the defective axles. Ball bushing axles may corrode and corrosion particle may cause contamination in the wells or the corrosion will cause movement issues and possibly spillage and causing contamination of the plate wells.",
      "recall_initiation_date": "20170105",
      "center_classification_date": "20170127",
      "termination_date": "20170824",
      "report_date": "20170208",
      "code_info": "N/A",
      "more_code_info": ""
    }
  ]
}