      "country": "United States",
      "city": "Andover",
      "address_1": "3000 Minuteman Rd",
      "reason_for_recall": "HeartStart XL+ battery charge time to 100% capacity at 35¿C (95¿F) is slightly longer than the specified 3 hour duration as stated in the labeling",
      "address_2": "",
      "product_quantity": "7157 units",
      "code_info": "Serial Numbers:  Serial Number Range  USO1100106-USO1100372  USN1100376-USN1100960  USD1100961-USD1101095  US11201095-US11201186  US21201187-US21201239  US31201240-US31201537  US41201538-US41201585  US51201586-US51201721  US61201722-US61201924  US71201925-US71202048  US81202049-US81202168  US91202169-US91202514  US01202515-US01202990  USN1202991-USN1203537  USD1203538-USD1203968  US11303969-US11303972  US21303980-US21304488  US31304489-US31305042  US41305043-US41305295  US51305296-US51305450  US61305451-US61306138  US71306139-US71306542  US81306543-US81306998  US91306999-US91307261  US01307306-US01307458",
      "center_classification_date": "20140226",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including Puerto Rico and the countries of  ALGERIA ANGOLA ARGENTINA AUSTRALIA AUSTRIA AZERBAIJAN BAHRAIN BELGIUM BOLIVIA BULGARIA CANADA CHILE CHINA COLOMBIA COSTA RICA CROATIA (local name: Hrvatska) CYPRUS CZECH REPUBLIC DENMARK DOMINICAN REPUBLIC EGYPT EL SALVADOR ESTONIA ETHIOPIA FINLAND FRANCE GABON GERMANY GREECE GUATEMALA HONG KONG HUNGARY ICELAND INDIA INDONESIA IRAN (ISLAMIC REPUBLIC OF) IRAQ IRELAND ISRAEL ITALY JORDAN KENYA KOREA, REPUBLIC OF KUWAIT LATVIA LEBANON LIBERIA LIBYAN ARAB JAMAHIRIYA LITHUANIA MACAU MALAYSIA MALTA MAURITIUS MOROCCO MYANMAR (Burma) NAMIBIA NETHERLANDS NEW ZEALAND NORWAY OMAN PALESTINE PERU PHILIPPINES POLAND PORTUGAL,QATAR ROMANIA RUSSIAN FEDERATION SAUDI ARABIA SENEGAL SERBIA SEYCHELLES SINGAPORE SLOVAKIA (Slovak Republic) SLOVENIA SOUTH AFRICA SPAIN SRI LANKA SWEDEN SWITZERLAND SYRIAN ARAB REPUBLIC TAIWAN, PROVINCE OF CHINA TANZANIA, UNITED REPUBLIC OF THAILAND TRINIDAD AND TOBAGO TUNISIA TURKEY UNITED ARAB EMIRATES UNITED KINGDOM and VIET NAM.",
      "state": "MA",
      "product_description": "Philips HeartStart XL+ Defibrillator/Monitor  Model: 861290    The HeartStart XL+ is intended for use in a hospital setting by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced life support or defibrillation. When operating as a semi-automatic external defibrillator in AED Mode, the HeartStart XL+ is suitable for use by medical personnel trained in basic life support that includes the use of an AED. When operating in Monitor, Manual Defibrillation or Pacing modes, the HeartStart XL+ is suitable for use by healthcare professionals trained in advance life support.",
      "report_date": "20140305",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Medical Systems, Inc.",
      "recall_number": "Z-1107-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67328",
      "termination_date": "20200903",
      "more_code_info": "",
      "recall_initiation_date": "20140124",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}