  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83994",
      "recalling_firm": "Intuitive Surgical, Inc.",
      "address_1": "1266 Kifer Rd Bldg 101",
      "address_2": "N/A",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US National Distribution   Alabama,  Arizona,  California,  Colorado,  Connecticut,  Delaware,  Florida,  Georgia,  Illinois,  Indiana,  Iowa,  Kentucky,  Louisiana,  Maryland,  Massachusetts,  Michigan,  Minnesota,  Missouri,  Montana,  Nebraska,  Nevada,  New Hampshire,  New Jersey,  New York,  North Carolina,  Ohio,  Oklahoma,  Oregon,  Pennsylvania,  Tennessee,  Texas,  Vermont,  Virginia,  Washington,  West Virginia,  Wisconsin,  Wyoming    OUS:None",
      "recall_number": "Z-1105-2020",
      "product_description": "Da Vinci Xi/X EndoWrist Monopolar Curved scissors (MCS), model number 470179-19, UDI: (00) 886874112298",
      "product_quantity": "286 Monopolar curved scissors",
      "reason_for_recall": "This recall is being initiated  because lots of EndoWrist Monopolar Curved Scissor instruments are not recognized by the integrated electrosurgical unit, which prohibits the activation of monopolar energy from the surgeon console.",
      "recall_initiation_date": "20181204",
      "center_classification_date": "20200206",
      "termination_date": "20220824",
      "report_date": "20200212",
      "code_info": "Serial Number/Manufacturing Lot: N11181003, N11181004"
    }
  ]
}