      "country": "United States",
      "city": "Mercedita",
      "address_1": "Turpeaux Industrial Park Rd 1 km 123.4",
      "reason_for_recall": "As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013.  The affected products are sterile.",
      "address_2": "",
      "product_quantity": "109",
      "code_info": "lot no:  62230449  62081271  62133855  62159156  62047629  62127760  61903333  62261628  61830478  61830479  62138082  62230452  61800380  61800381",
      "center_classification_date": "20160312",
      "distribution_pattern": "US Nationwide Distribution",
      "state": "PR",
      "product_description": "Product 7 consists of all products under  product code HWC, and same usage:  Item no:\t  47234801435\t3.5MM CORT SCR X 14MM  47234801635\t3.5MM CORT SCR X 16MM  47234806535\t3.5MM CORT SCR X 65MM  47234807035\t3.5MM CORT SCR X 70MM  47234807535\t3.5MM CORT SCR X 75MM  47234808035\t3.5MM CORT SCR X 80MM  47234808535\t3.5MM CORT SCR X 85MM  47234809035\t3.5MM CORT SCR X 90MM  Product Usage:  A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.",
      "report_date": "20160323",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Manufacturing B.V.",
      "recall_number": "Z-1105-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "72900",
      "termination_date": "20170413",
      "more_code_info": "",
      "recall_initiation_date": "20160111",
      "postal_code": "00715",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}