  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Omaha",
      "address_1": "PO Box 27261",
      "reason_for_recall": "Product resistance has fallen outside of label claim within expiry.",
      "address_2": "8607 Park Dr",
      "product_quantity": "328/ 50-count/ boxes; 16,400 individual units",
      "code_info": "Lot # 637, exp. date 12/2016",
      "center_classification_date": "20170126",
      "distribution_pattern": "Product was distributed to:  AL, AR, CA, FL, IA, MO, NE, NC, NY, OH, and WA.  The product was also distributed internationally, including Mexico and Canada.  There was no goverment or military distribution.",
      "state": "NE",
      "product_description": "ProSpore 4 mL Ampoule Biological indicator for steam sterilization, contains Geobacillus stearothermophilus, Catalog Number: PS-6-50, contains 50 Ampoules/Box.      Product Usage:  Biological indicator for steam sterilization at 121 deg C intended for use by a health care provider to accompany products being sterilized through steam sterilization procedure and to monitor adequacy of sterilization.",
      "report_date": "20170201",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Mesa Laboratories Inc",
      "recall_number": "Z-1104-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "76101",
      "termination_date": "20170126",
      "recall_initiation_date": "20160404",
      "postal_code": "68127-3623",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}