  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91291",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "FL, GA, IN, MD, MS, NY, OH, PA, RI, SC, VA",
      "recall_number": "Z-1103-2023",
      "product_description": "SPiN Thoracic Navigation System  Model: SYS-4000 (SYS-4000D)",
      "product_quantity": "13 units",
      "reason_for_recall": "A grainy image is observed when using a SPiN Vision Single-Use Flexible Bronchoscope (models INS-7100 or INS-7130) with the SYS-4000 system due to an incorrect DVI cable supplied may result in prolonged  procedure/surgery due to additional time to  troubleshoot the device",
      "recall_initiation_date": "20221213",
      "center_classification_date": "20230207",
      "report_date": "20230215",
      "code_info": "UDI: 00815686020101 Serial Numbers: 85330200430, 87896220519, 87899220524, 87379210921, 87835220505, 87832220502, 87767220420, 87777220415, 87827220427, 87736220412, 87891220517, 87864220509, 87738220412",
      "more_code_info": ""
    }
  ]
}