      "status": "Ongoing",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91526",
      "recalling_firm": "Wishbone Medical, Inc.",
      "address_1": "100 Capital Dr",
      "address_2": "",
      "postal_code": "46582-6704",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the state of South Carolina.",
      "recall_number": "Z-1102-2023",
      "product_description": "WishBone Medical Smart Correction Dual Joint Express Struts, external skeleton fixation device    Type         Part Number  XXS           10-702-080095-0  XS              10-702-095115-0  Short        10-702-115150-0  Medium    10-702-140200  Long          10-702-190300-0",
      "product_quantity": "195",
      "reason_for_recall": "Firm identified that user may be unable to adjust the length of the Smart Correction Dual Joint Express Struts due to the collar release pin backing out. This may lead to a delay in correction.",
      "recall_initiation_date": "20230104",
      "center_classification_date": "20230207",
      "report_date": "20230215",
      "code_info": "XXS, Part No. 10-702-080095-0 UDI-DI B5621070208009500 Lot 1048176, 1048221  XS, Part No. 10-702-095115-0 UDI-DI B5621070209511500 Lot 1048177  Short, Part No. 10-702-115150-0 UDI-DI B5621070211515000 Lot 1048178, 1048222, 1048223  Medium, Part No. 10-702-140200 UDI-DI B562107021402000 Lot 1048179  Long, Part No.  10-702-190300-0 UDI-DI B5621070219030000 Lot 1048180",
      "more_code_info": ""
    }
  ]
}