{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Crumlin (North)",
      "state": "N/A",
      "country": "Ireland",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84547",
      "recalling_firm": "Randox Laboratories Ltd.",
      "address_1": "55 Diamond Road",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: NY and DE    OUS: France, Lithuania, Poland, Spain, Switzerland, Turkey, Australia, Hong Kong, Bosnia and Herzegovina, Costa Rica, Ecuador, Gambia, New Zealand and Solomon Islands.",
      "recall_number": "Z-1100-2020",
      "product_description": "Liquid Clinical Chemistry Control, Catalogue Number: LAE4217, Size: 12 x 5 ml",
      "product_quantity": "2",
      "reason_for_recall": "A transcription error has occurred for the Randox Lipase colorimetric value in Liquid Clinical Chemistry Control lot 1021 UE. The target and range are listed as 50 U/L (40 - 60) and should be 72U/L (58 - 86).",
      "recall_initiation_date": "20191213",
      "center_classification_date": "20200206",
      "termination_date": "20201002",
      "report_date": "20200212",
      "code_info": "Catalogue Number: LAE4217  GTIN: 05055273208993  Lot Number: 1021UE (EXP: 2020-06-28)"
    }
  ]
}