  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Waukesha",
      "address_1": "3000 N Grandview Blvd",
      "reason_for_recall": "GE Healthcare has recently become aware of a potential issue due to image flip in frequency encoding direction associated with a specific workflow of your Discovery MR750w system.  Image flip issue in frequency encoding direction (including S/I, A/P and R/L) can result in incorrect localization of  anatomy and pathological findings.",
      "address_2": "",
      "product_quantity": "46",
      "code_info": "Serial Number  00000000UA0027  00000000UA0012  00000000UA0029  00000000UA0031  00000000UA0030  00000000UA0028  00000000UA0034  00000000UA0038  00000000UA0048  00000000UA0046  00000000UA0052  00000000UA0041  00000000UA0011A  00000000UA0055  00000000UA0056  00000000UA0057  00000000UA0064  00000000UA0053  00000000UA0044  00000000UA0047  00000000UA0010  00000000UA0013  00000000UA0017  00000000UA0042  00000000UA0016  00000000UA0019  00000000UA0015  00000000UA0058  00000000UA0051  00000000UA0025  00000000UA0024  00000000UA0020  00000000UA0040  00000000UA0018  00000000UA0036  00000000UA0043  00000000UA0054  00000000UA0022  00000000UA0023  00000000UA0045  00000000UA0039  00000000UA0035  00000000UA0049  00000000UA0021  00000000UA0033  00000000UA0037",
      "center_classification_date": "20130411",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including the states of:  AZ, CA, CO, FL, IL, MD, NJ, NY, OH, OK, TX. and the countries of Australia, Belarus, Canada,  France, Germany, India, Japan, Lebanon, New Zealand, Saudi Arabia, Singapore, Spain, Sweden, Switzerland and United Kingdom.",
      "state": "WI",
      "product_description": "GE Healthcare, Discovery MR750w 3.0T System.   Part Number: 5352293.    Product Usage:  It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra,dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate,blood vessels, and musculoskeletal regions of the body.",
      "report_date": "20130417",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GE Healthcare, LLC",
      "recall_number": "Z-1100-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64763",
      "termination_date": "20140314",
      "more_code_info": "",
      "recall_initiation_date": "20120423",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}