  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cambridge",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87090",
      "recalling_firm": "Cytocell Ltd.",
      "address_1": "418 Cambridge Science Park, Milton Road",
      "address_2": "Oxford Gene Technology",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Domestic distribution to California, Illinois, Michigan, Minnesota, New Jersey, Texas, and Washington. Foreign distribution to Canada.",
      "recall_number": "Z-1099-2021",
      "product_description": "Aquarius LPH 511-A CRLF2 Distal Probe Green, model no. LPH 511-A    The CRLF2 Distal probe, labelled in green, consists of two probes (71kb and 131kb) that are situated distal (telomeric)  to the CRLF2 gene",
      "product_quantity": "72",
      "reason_for_recall": "Users may observe very faint additional locus specific signals at 11q13. There is a low risk of a normal sample being reported as not normal in a resulting laboratory developed test (LDT) that utilizes the LPH511-A CRLF2 Distal Probe Green.",
      "recall_initiation_date": "20201222",
      "center_classification_date": "20210208",
      "termination_date": "20220607",
      "report_date": "20210217",
      "code_info": "Lots 072240, 071390, and 072693"
    }
  ]
}