{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Portage",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84721",
      "recalling_firm": "Stryker Instruments Div. of Stryker Corporation",
      "address_1": "4100 E Milham Ave",
      "address_2": "N/A",
      "postal_code": "49002-9704",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution nationwide. Foreign distribution to Netherlands and Australia.",
      "recall_number": "Z-1099-2020",
      "product_description": "Telescopic Uncoated (Push  Button), Product No. SHK TSPL. The product is a Bovie pencil that is 75 inches long compressed, with a telescoping shaft. UDI GTIN 37613327466332",
      "product_quantity": "15150",
      "reason_for_recall": "Hairline fractures in a component could allow electrical current to arc out of the device, resulting in a burn to the patient and/or user.",
      "recall_initiation_date": "20200113",
      "center_classification_date": "20200206",
      "termination_date": "20210810",
      "report_date": "20200212",
      "code_info": "1922042, 1924003, 1929040,  1929048"
    }
  ]
}