  "results": [
    {
      "status": "Terminated",
      "city": "Arvada",
      "state": "CO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82365",
      "recalling_firm": "LivaNova USA Inc.",
      "address_1": "14401 W 65th Way",
      "address_2": "N/A",
      "postal_code": "80004-3503",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "US Distribution to states of: IN, MO, and OH.",
      "recall_number": "Z-1099-2019",
      "product_description": "SORIN Heater-Cooler System 3T, REF 16-02-85    Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung machine featuring a separate temperature control for extracorporeal perfusion of durations of up to 6 hours.",
      "product_quantity": "6 devices",
      "reason_for_recall": "For six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM).",
      "recall_initiation_date": "20180914",
      "center_classification_date": "20190405",
      "termination_date": "20220914",
      "report_date": "20190417",
      "code_info": "Serial numbers 16S13187, 16S14887, 16S15335, 16S15336, 16S15337, and 16S11877."
    }
  ]
}