      "status": "Terminated",
      "city": "Portage",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84721",
      "recalling_firm": "Stryker Instruments Div. of Stryker Corporation",
      "address_1": "4100 E Milham Ave",
      "address_2": "N/A",
      "postal_code": "49002-9704",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution nationwide. Foreign distribution to Netherlands and Australia.",
      "recall_number": "Z-1098-2020",
      "product_description": "Telescopic Smoke Evac Pencil,  PB, Coated, Product No. SHKTSPCL. The product is a Bovie pencil that is 75 inches long compressed, with a telescoping shaft. UDI GTIN 37613327466332",
      "product_quantity": "15150",
      "reason_for_recall": "Hairline fractures in a component could allow electrical current to arc out of the device, resulting in a burn to the patient and/or user.",
      "recall_initiation_date": "20200113",
      "center_classification_date": "20200206",
      "termination_date": "20210810",
      "report_date": "20200212",
      "code_info": "1923011, 1923050, 1924004,  1924005, 1924015, 1929038,  1929039, 1929055"
    }
  ]
}