  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Clemente",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82291",
      "recalling_firm": "ICU Medical, Inc.",
      "address_1": "951 Calle Amanecer",
      "address_2": "N/A",
      "postal_code": "92673-6212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution: US nationwide and countries of: Canada, France, and Singapore.",
      "recall_number": "Z-1097-2019",
      "product_description": "icumedical ChemoLock Closed Vial Spike w/Skirt,   (a) REF CL-80S-4 (4 units)  (b) REF CL-80S-5 (5 units).     Used in IV administration sets.     The ChemoLock Closed System Drug Transfer Device prevents the transfer of environmental contaminants, including bacterial and airborne contaminants into the system, and the escape of drug or vapor concentrations outside the system. The ChemoLock is needlefree and cannot be deactivated, which will passively aid in preventing needlestick injuries and the exposure to cytotoxic medications for healthcare personnel.",
      "product_quantity": "23800 devices",
      "reason_for_recall": "There is a potential for certain lots of ChemoLock and ChemoClave Vial Spikes to have plastic burr particulate originating from the protective cap used in the assembly of the device.",
      "recall_initiation_date": "20190221",
      "center_classification_date": "20190415",
      "termination_date": "20240604",
      "report_date": "20190424",
      "code_info": "(a) REF CL-80S-4, Lot Numbers:  3820520, 3820521, 3844785    (b) REF CL-80S-5, Lot Numbers:  3820516, 3820523, 3839238, 3839239, 3844781, 3844782, 3853998, 3853999, 3866769, 3866770, 3872090, 3910645, 3932908, 3932910, 3946770, 3951236"
    }
  ]
}