      "reason_for_recall": "GE Healthcare has become aware of a potential safety issue due to the S4-10-D Ultrasound Probe used in conjunction with the LOGIQ S8 and LOGIQ S7 Ultrasound systems. It was also discovered that this issue also leads to a specification issue of the acoustic output of this probe. There is a potential for a skin burn when the flow model in the cardiac application is activated with the S4-10-D Ultraso",
      "address_2": "",
      "product_quantity": "2061",
      "code_info": "Mfg Lot or Serial #  176132SU9  176138SU6  178028SU7  158247SU7  172656SU1  187599SU6  201614SU5  201615SU2  195234SU0  195237SU3  161686SU1  203123SU5  170272SU9  189160SU5  189162SU1  170273SU7  185619SU4  176924SU9  191894SU5  159964SU6  159768SU1  170197SU8  158248SU5  171540SU8  171541SU6  160741SU5  160742SU3  160743SU1  159766SU5  197761SU0  198949SU0  195241SU5  189928SU5  164268SU5  159963SU8  185116SU1  195242SU3  201967SU7  166218SU8  185968SU5  189165SU4  189166SU2  189167SU0  189168SU8  189169SU6  195235SU7  195236SU5  171371SU8  168779SU7  189163SU9  190393SU9  159767SU3  167274SU0  201464SU5  195233SU2  171369SU2  177085SU8  186501SU3  171370SU0  187022SU9  191063SU7  185969SU3  168788SU8  159965SU3  201612SU9  159604SU8  178592SU2  178593SU0  178594SU8  178595SU5  185620SU2  185621SU0  176926SU4  176925SU6  191893SU7  177084SU1  192952SU0  195238SU1  195239SU9  195240SU7  201971SU9  201972SU7  202046SU9  195881SU8  186579SU9  204275SU2  171542SU4  171947SU5  192951SU2  201906SU5  204921SU1  204922SU9  204923SU7  204924SU5  204925SU2  189161SU3  181434SU2  158768SU2  183107SU2  166181SU8  169509SU7  170671SU2  187015SU3  187014SU6  183118SU9  170672SU0  176130SU3  182380SU6  188131SU7  188121SU8  159776SU4  159792SU1  191471SU2  182223SU8  182222SU0  183104SU9  166172SU7  197249SU6  194194SU7  168080SU0  195876SU8  198407SU9  190385SU5  190754SU2  190390SU5  182654SU4  182655SU1  182656SU9  189924SU4  176566SU8  176567SU6  190752SU6  190753SU4  194561SU7  194562SU5  194563SU3  194564SU1  194565SU8  196258SU8  183907SU5  200715SU1  188545SU8  176939SU7  190505SU8  190506SU6  190507SU4  192947SU0  180973SU0  183102SU3  176938SU9  185798SU6  200709SU4  180363SU4  183660SU0  183661SU8  189354SU4  202264SU8  192059SU4  192060SU2  192116SU2  192115SU4  193638SU4  199071SU2  192120SU4  183470SU4  183471SU2  183472SU0  183473SU8  186887SU6  192058SU6  192061SU0  192062SU8  192063SU6  192065SU1  192105SU5  192110SU5  192114SU7  192119SU6  192546SU0  192547SU8  192548SU6  192549SU4  192550SU2  192551SU0  192552SU8  192553SU6  193342SU3  194554SU2  196884SU1  196885SU8  197658SU8  197760SU2  198879SU9  199237SU9  199238SU7  199239SU5  200704SU5  200706SU0  200717SU7  201237SU5  201238SU3  201239SU1  201616SU0  201617SU8  201907SU3  201965SU1  201966SU9  202262SU2  202263SU0  202265SU5  202266SU3  202267SU1  202494SU1  202495SU8  202496SU6  202497SU4  202498SU2  202725SU8  202732SU4  203033SU6  192118SU8  192064SU4  200173SU3  192113SU9  192112SU1  192106SU3  194555SU9  192117SU0  192107SU1  192111SU3  200409SU1  193344SU9  182653SU6  194542SU7  190400SU2  190396SU2  190380SU6  190391SU3  194537SU7  194540SU1  194538SU5  190395SU4  190399SU6  190387SU1  190384SU8  190388SU9  190382SU2  189691SU9  183912SU5  159246SU8  198930SU0  198931SU8  198932SU6  189693SU5  191015SU7  194182SU2  187011SU2  191016SU5  193903SU2  158805SU2  180762SU7  197254SU6  170669SU6  198929SU2  183920SU8  170679SU5  170683SU7  158802SU9  162595SU3  185297SU9  194190SU5  182379SU8  176132SU9  183108SU0  168063SU6  168064SU4  169487SU6  169506SU3  191567SU7  192568SU4  176121SU2  158806SU0  195863SU6  195864SU4  195865SU1  195872SU7  201236SU7  183922SU4  185294SU6  187597SU0  159784SU8  186590SU6  196272SU9  185302SU7  185305SU0  188124SU2  165587SU7  168058SU6  191562SU8  170676SU1  198947SU4  162594SU6  163224SU9  198924SU3  198926SU8  159789SU7  191563SU6  198940SU9  198941SU7  183913SU3  183914SU1  198910SU2  198911SU0  185301SU9  198945SU8  182227SU9  178022SU0  178023SU8  198424SU4  158800SU3  191564SU4  165585SU1  168056SU0  191013SU2  182228SU7  182376SU4  191560SU2  198946SU6  183105SU6  197256SU1  187598SU8  168068SU5  170675SU3  198921SU9  185307SU6  165588SU5  175622SU0  175623SU8  168070SU1  180761SU9  182382SU2  191014SU0  201235SU9  176136SU0  189175SU3  198922SU7  198923SU5  159782SU2  159783SU0  168057SU8  182230SU3  183106SU4  185299SU5  185304SU3  182378SU0  187596SU2  198919SU3  198920SU1  168060SU2  198925SU0  185295SU3  198943SU3  198944SU1  163225SU6  183109SU8  183110SU6  198939SU1  168059SU4  188128SU3  185300SU1  166185SU9  185306SU8  178018SU8  183918SU2  183921SU6  170674SU6  201444SU7  161687SU9  161688SU7  161689SU5  161690SU3  161691SU1  161692SU9  189176SU1  198942SU5  189172SU0  159788SU9  187594SU7  187595SU4  197255SU3  189173SU8  198915SU1  165584SU4  176135SU2  201439SU7  165583SU6  165586SU9  182377SU2  162593SU8  191561SU0  196259SU6  198948SU2  195871SU9  198099SU4  199010SU0  168073SU5  168074SU3  168075SU0  168071SU9  159778SU0  175624SU6  187591SU3  191544SU6  178027SU9  181605SU7  166178SU4  194528SU6  183120SU5  178026SU1  171946SU7  182224SU6  170897SU3  198897SU1  178016SU2  183915SU8  194535SU1  199497SU9  169502SU2  169508SU9  185308SU4  186451SU1  178029SU5  158763SU3  180768SU4  161694SU5  176138SU6  159249SU2  159249SU2  183116SU3  159775SU6  162598SU7  190405SU1  190406SU9  171944SU2  185292SU0  163230SU6  163223SU1  158759SU1  158766SU6  175617SU0  196271SU1  183911SU7  194187SU1  196268SU7  193954SU5  166176SU8  166177SU6  166174SU3  168055SU2  170673SU8  166180SU0  187018SU7  185303SU5  189180SU3  159247SU6  191466SU2  180763SU5  194186SU3  191543SU8  183113SU0  183114SU8  183115SU5  201441SU3  168838SU1  170898SU1  158767SU4  162599SU5  201443SU9  188120SU0  159779SU8  191558SU6  169511SU3  178038SU6  166183SU4  196264SU6  191549SU5  182221SU2  201446SU2  168072SU7  184258SU2  186588SU0  194526SU0  159781SU4  169505SU5  190410SU1  176128SU7  159785SU5  159786SU3  178019SU6  178020SU4  192954SU6  192956SU1  194544SU3  201442SU1  162597SU9  196263SU8  169493SU4  194534SU4  159793SU9  178017SU0  183910SU9  169510SU5  185291SU2  171945SU9  193960SU2  190401SU0  192955SU3  169513SU9  178025SU3  191554SU5  168076SU8  165595SU0  187008SU8  187588SU9  191550SU3  159777SU2  169503SU0  176131SU1  169504SU8  193904SU0  183888SU7  166179SU2  163229SU8  198933SU4  163231SU4  180765SU0  199496SU1  189183SU7  181433SU4  193963SU6  159774SU9  162596SU1  166175SU0  196266SU1  191465SU4  191679SU0  193964SU4  187021SU1  194527SU8  198427SU7  188119SU2  183926SU5  190403SU6  190404SU4  197259SU5  198436SU8  191463SU9  197169SU6  197240SU5  198101SU8  197173SU8  195858SU6  197242SU1  198323SU8  197172SU0  198317SU0  197170SU4  198322SU0  197243SU9  190381SU4  201226SU8  198100SU0  178028SU7  198321SU2  195875SU0  198316SU2  198406SU1  198098SU6  191545SU3  172890SU6  198928SU4  183119SU7  161695SU2  176015SU6  158764SU1  189674SU5  193952SU9  170681SU1  184445SU5  189440SU1  189441SU9  202470SU1  202492SU5  202493SU3  179387SU6  179388SU4  200718SU5  165598SU4  176013SU1  159248SU4  191547SU9  186399SU2  159244SU3  183908SU3  158807SU8  165599SU2  159251SU8  183916SU6  190407SU7  181431SU8  182381SU4  176122SU0  165590SU1  183917SU4  165589SU3  193900SU8  188132SU5  165600SU8  170894SU0  191546SU1  158760SU9  169501SU4  187010SU4  178021SU2  176127SU9  198934SU2  187585SU5  187012SU0  165591SU9  165592SU7  165593SU5  165594SU3  170895SU7  193902SU4  194181SU4  171942SU6  175618SU8  168053SU7  187013SU8  170896SU5  183924SU0  183925SU7  198408SU7  199014SU2  198405SU3  183112SU2  197171SU2  159951SU3  192701SU1  153258SU9  195256SU3  196255SU4  192122SU0  192121SU2  192706SU0  193643SU4  181309SU6  153255SU5  159953SU9  174785SU6  180006SU9  159956SU2  175346SU6  159241SU9  164232SU1  196256SU2  192702SU9  159601SU4  158245SU1  171935SU0  196881SU7  196882SU5  168784SU7  188593SU8  188594SU6  188595SU3  176444SU8  168782SU1  176443SU0  188596SU1  162831SU2  196253SU9  200712SU8  159954SU7  195713SU3  188598SU7  192125SU3  172855SU9  176440SU6  190755SU9  199241SU1  159957SU0  175348SU2  176445SU5  188591SU2  164237SU0  162830SU4  180365SU9  192705SU2  180005SU1  181308SU8  181308SU8  199076SU1  159948SU9  159949SU7  173818SU6  176447SU1  164238SU8  195710SU9  188597SU9  175349SU0  153260SU5  153254SU8  193639SU2  176441SU4  192127SU9  195209SU2  166279SU0  167283SU1  166275SU8  167282SU3  166283SU2  167285SU6  166276SU6  166278SU2  167277SU3  167284SU9  167286SU4  167278SU1  166280SU8  166281SU6  166282SU4  167279SU9  167280SU7  167281SU5  166277SU4  164233SU9  175347SU4  182385SU5  182435SU8  175350SU8  181312SU0  180000SU2  162832SU0  200711SU0  159608SU9  195712SU5  180366SU7  195259SU7  168787SU0  192703SU7  168786SU2  174784SU9  193641SU8  193642SU6  164239SU6  164240SU4  174781SU5  153256SU3  196878SU3  196880SU9  195711SU7  195255SU5  168783SU9  199072SU0  159602SU2  159605SU5  181314SU6  164236SU2  190758SU3  190756SU7  164235SU4  168781SU3  199244SU5  190757SU5  164234SU7  196275SU2  196276SU0  196277SU8  196278SU6  183468SU8  200340SU8  201465SU2  195714SU1  180004SU4  171934SU3  178952SU8  171933SU5  159947SU1  196257SU0  171931SU9  178954SU4  199075SU3  188592SU0  174782SU3  174783SU1  176442SU2  158244SU4  192707SU8  200342SU4  200714SU4  153257SU1  176446SU3  168780SU5  183469SU6  193640SU0  195257SU1  159950SU5  159603SU0  159596SU6  178953SU6  182388SU9  180367SU5  182384SU8  159952SU1  171936SU8  172853SU4  172854SU2  173816SU0  173817SU8  190759SU1  182436SU6  159600SU6  179999SU8  181313SU8  200344SU0  153259SU7  159599SU0  159607SU1  159955SU4  159959SU6  159960SU4  159961SU2  162833SU8  162834SU6  171932SU7  173819SU4  180001SU0  180002SU8  180003SU6  181310SU4  182233SU7  182234SU5  182235SU2  182386SU3  182387SU1  192126SU1  192128SU7  192129SU5  192130SU3  192131SU1  192132SU9  192133SU7  192704SU5  195208SU4  195210SU0  195212SU6  195213SU4  195214SU2  195215SU9  195216SU7  195217SU5  195218SU3  195219SU1  195220SU9  195221SU7  195222SU5  195223SU3  195223SU3  195258SU9  196254SU7  196274SU5  196879SU1  202259SU8  203284SU5  159597SU4  159962SU0  158243SU6  159958SU8  159242SU7  164241SU2  168785SU4  192123SU8  192124SU6  195211SU8  199242SU9  203280SU3  203282SU9  203283SU7  159598SU2  159606SU3  180364SU2  199073SU8  200713SU6  202258SU0  202260SU6  202261SU4  204265SU3  204266SU1  204267SU9  204268SU7  203281SU1  199240SU3  199243SU7  200341SU6  200343SU2  201072SU6  201073SU4  201074SU2  201075SU9  201076SU7  201973SU5  201974SU3  201975SU0  201976SU8  201977SU6  202257SU2  199074SU6  185457SU9  193385SU2  193386SU0  193387SU8  197759SU4  200710SU2  199606SU5  199789SU9  200679SU9  200680SU7  200960SU3  200961SU1  201077SU5  201214SU4  201610SU3  201613SU7  201618SU6  201969SU3  202467SU7  202468SU5  168065SU1  168066SU9  183103SU1  182236SU0  167276SU5  170347SU9  191696SU4  193657SU4  194546SU8  199357SU5  200678SU1  200708SU6  203046SU8  160385SU1  184516SU3  187647SU3  161693SU7  187020SU3  176017SU2  187019SU5  165582SU8  168319SU2  197942SU6  186622SU7  184408SU3  190751SU8  180773SU4  180774SU2  189164SU7  180771SU8  180772SU6  170666SU2  170668SU8  170667SU0  179252SU2  171948SU3  179458SU5  181425SU0  182229SU5  170677SU9  183919SU0  191472SU0  189171SU2  182374SU9  197245SU4  197246SU2  191540SU4  178030SU3  158761SU7  176141SU0  169512SU1  162600SU1  198439SU2  166184SU2  194539SU3  187592SU1  195874SU3  196265SU3  195866SU9  200705SU2  193474SU4  159243SU5  176126SU1  182225SU3  194185SU5  194191SU3  195857SU8  191541SU2  171938SU4  190392SU1  166173SU5  158765SU8  183111SU4  190409SU3  195867SU7  159250SU0  176124SU6  188123SU4  170680SU3  176016SU4  176014SU9  167275SU7  170271SU1  170198SU6  200707SU8  189675SU2  158259SU2  159240SU1  193357SU1  159245SU0  176134SU5  193449SU6  193450SU4  198893SU0  169507SU1  193949SU5  203456SU9  167648SU5  168521SU3  179023SU7  179324SU9  179325SU6  179326SU4  191243SU5  188357SU8  193343SU1  193637SU6  176887SU8  195232SU4  203337SU1  158246SU9  160531SU0  203458SU5  179997SU2  191892SU9  203450SU2  192109SU7  171032SU6  186601SU1  179998SU0  185067SU6  193038SU7  186726SU6  203125SU0  203338SU9  169486SU8  183101SU5  185315SU9  160530SU2  185799SU4  185800SU0  171033SU4  168778SU9  187782SU8  171255SU3  195229SU0  195230SU8  195231SU6  186727SU4  192108SU9  193338SU1  193339SU9  193340SU7  193341SU5  202469SU3  203459SU3  170682SU9  171939SU2  169491SU8  180764SU3  169492SU6  198420SU2  181430SU0  169490SU0  185298SU7  158801SU1  165597SU6  196267SU9  162592SU0  163227SU2  170678SU7  169489SU2  178024SU6  198909SU4  191559SU4  196260SU4  196261SU2  158762SU5  194532SU8  187584SU8  166182SU6  183117SU1  201234SU2  168067SU7  189170SU4  169488SU4  165581SU0  193957SU8  163226SU4  168054SU5  158803SU7  195862SU8  195861SU0  198876SU5  198877SU3  186623SU5  179272SU0  193947SU9  198875SU7  200164SU2  200165SU9  191065SU2  191064SU5  172241SU2  184017SU2  184517SU1  189940SU0  192544SU5  192545SU2  172240SU4  168854SU8  168855SU5  168079SU2  168077SU6  168078SU4  190373SU1  190383SU0  187590SU5  165596SU8  180767SU6  182375SU6  163228SU0  190414SU3  170892SU4  171940SU0  159252SU6  193950SU3  166186SU7  194192SU1  194533SU6  193448SU8  194193SU9  183923SU2  198883SU1  191568SU5  187009SU6  169498SU3  169499SU1  191462SU1  198912SU8  198913SU6  176008SU1  176009SU9  176010SU7  176011SU5  176012SU3  188122SU6  176140SU2  169500SU6  176142SU8  176143SU6  169494SU2  169495SU9  168061SU0  168062SU8  158804SU5  159791SU3  182383SU0  186589SU8  189174SU6  181427SU6  159790SU5  176137SU8  176145SU1  176146SU9  178014SU7  190377SU2  198884SU9  198892SU2  198917SU7  198918SU5  187593SU9  176119SU6  176120SU4  188125SU9  178015SU4  189177SU9  190374SU9  198886SU4  176118SU8  176144SU4  169496SU7  169497SU5  175619SU6  175620SU4  175621SU2  181428SU4  181429SU2  190376SU4  198888SU0  181426SU8  191461SU3  190413SU5  190411SU9  190412SU7  200956SU1  193882SU8  179457SU7  160126SU9  189429SU4  193879SU4  200952SU0  191536SU2  191538SU8  200957SU9  202730SU8  186631SU8  182659SU3  182660SU1  171022SU7  171031SU8  169348SU0  185137SU7  185140SU1  184029SU7  189439SU3  196893SU2  189433SU6  193887SU7  159771SU5  159133SU8  159256SU7  159135SU3  158236SU0  185127SU8  184041SU2  159773SU1  158237SU8  184032SU1  185134SU4  181235SU3  193135SU1  196877SU5  198900SU3  185135SU1  189936SU8  197659SU6  188974SU0  181642SU0  185138SU5  193497SU5  193500SU6  188973SU2  159772SU3  198414SU5  200955SU3  181643SU8  169788SU7  184025SU5  184027SU1  184028SU9  197178SU7  202727SU4  193136SU9  165329SU4  184022SU2  197660SU4  184021SU4  178009SU7  203039SU3  165576SU0  165579SU4  181644SU6  182437SU4  193884SU4  169784SU6  158239SU4  164379SU0  164381SU6  193137SU7  190421SU8  182440SU8  180109SU1  193138SU5  193141SU9  197174SU6  191534SU7  192309SU3  184039SU6  176004SU0  169346SU4  193134SU4  173822SU8  173823SU6  193892SU7  193898SU4  169785SU3  171030SU0  178951SU0  195243SU1  188972SU4  184030SU5  186634SU2  164265SU1  164380SU8  159134SU6  203901SU4  190423SU4  203906SU3  169780SU4  181645SU3  186624SU3  190416SU8  190417SU6  185126SU0  176005SU7  179455SU1  200959SU5  160528SU6  197180SU3  198905SU2  179452SU8  185133SU6  193498SU3  193501SU4  192313SU5  181641SU2  177276SU3  169781SU2  189931SU9  159254SU2  193883SU6  165746SU9  184020SU6  189432SU8  193132SU8  196888SU2  176007SU3  164377SU4  185125SU2  192312SU7  193139SU3  193140SU1  200954SU6  198412SU9  186629SU2  169782SU0  203041SU9  169343SU1  186626SU8  192310SU1  165325SU2  165326SU0  165327SU8  165332SU8  180107SU5  198410SU3  189932SU7  189938SU4  198417SU8  195252SU2  198413SU7  184019SU8  184036SU2  185123SU7  193899SU2  197181SU1  191535SU4  197179SU5  196892SU4  197183SU7  188976SU5  158235SU2  158242SU8  160529SU4  189430SU2  185131SU0  189431SU0  203043SU5  186637SU5  159770SU7  185121SU1  196894SU0  164267SU7  185118SU7  186632SU6  184034SU7  189428SU6  193133SU6  182662SU7  193894SU3  165577SU8  171024SU3  189437SU7  189438SU5  186633SU4  185122SU9  159769SU9  173363SU3  169345SU6  169344SU9  186630SU0  169347SU2  171023SU5  198908SU6  202729SU0  203042SU7  159257SU5  173821SU0  203036SU9  169777SU0  169787SU9  182441SU6  194548SU4  193142SU7  177277SU1  196887SU4  197182SU9  164378SU2  177081SU7  175345SU8  175344SU1  197184SU5  178008SU9  171027SU6  197177SU9  196891SU6  200958SU7  195245SU6  195254SU8  171029SU2  172651SU2  182658SU5  177275SU5  203040SU1  158238SU6  189933SU5  186635SU9  190418SU4  184026SU3  184040SU4  177083SU3  198418SU6  195250SU6  172649SU6  184031SU3  198906SU0  182442SU4  197185SU2  164264SU4  165323SU7  165324SU5  164263SU6  164266SU9  186625SU0  185117SU9  186628SU4  185120SU3  203034SU4  196889SU0  197176SU1  186636SU7  189435SU1  189436SU9  193876SU0  193878SU6  193880SU2  193885SU1  191539SU6  182439SU0  159253SU4  160128SU5  184023SU0  169778SU8  169786SU1  171028SU4  182438SU2  184033SU9  192311SU9  195244SU9  175342SU5  203038SU5  182657SU7  196886SU6  185136SU9  189929SU3  185119SU5  172648SU8  172650SU4  178007SU1  165328SU6  165330SU2  198411SU1  198902SU9  185128SU6  185132SU8  189934SU3  193881SU0  171025SU0  202726SU6  190420SU0  179456SU9  194553SU4  197944SU2  171021SU9  203035SU1  193891SU9  176006SU5  165578SU6  177082SU5  179454SU4  200953SU8  169783SU8  172652SU0  184038SU8  186627SU6  191537SU0  180105SU9  175343SU3  184035SU4  184037SU0  190422SU6  189935SU0  190415SU0  160130SU1  198904SU5  196890SU8  196897SU3  194552SU6  172654SU6  189434SU4  193499SU1  198901SU1  173365SU8  189937SU6  188975SU7  179453SU6  193895SU0  168692SU2  165749SU3  166467SU1  165739SU4  165747SU7  165745SU1  165744SU4  166465SU5  166468SU9  165738SU6  168691SU4  165741SU0  165743SU6  165740SU2  165748SU5  165742SU8  166464SU8  172655SU3  198907SU8  185124SU5  190424SU2  181646SU1  173820SU2  196896SU5  184024SU8  177278SU9  184018SU0  160527SU8  178006SU3  195879SU2  158240SU2  158241SU0  159132SU0  160127SU7  160129SU3  165331SU0  165333SU6  169779SU6  171026SU8  193888SU5  193889SU3  193890SU1  193893SU5  193896SU8  193897SU6  194551SU8  199487SU0  159255SU9  165580SU2  189939SU2  190419SU2  202728SU2  197175SU3  158235SU2  158238SU6  158239SU4  195877SU6  196895SU7  198415SU2  198416SU0  198419SU4  180108SU3  180110SU9  180632SU2  181231SU2  181232SU0  181233SU8  181234SU6  185139SU3  188977SU3  189930SU1  193877SU8  193886SU9  194547SU6  194549SU2  194550SU0  195246SU4  195247SU2  195248SU0  195249SU8  195251SU4  195253SU0  195878SU4  195880SU0  198903SU7  199353SU4  199354SU2  199355SU9  199356SU7  201070SU0  201071SU8  202722SU5  202723SU3  203037SU7  201619SU4  185130SU2  187583SU0  193143SU5  153251SU4  153252SU2  153253SU0  158799SU7  158808SU6  159780SU6  159787SU1  162601SU9  163222SU3  170670SU4  170893SU2  171937SU6  171941SU8  171943SU4  172243SU8  172653SU8  173362SU5  173364SU1  176123SU8  176125SU3  176129SU5  176260SU8  177080SU9  180108SU3  180766SU8  181432SU6  182226SU1  182661SU9  183909SU1  185129SU4  185293SU8  185296SU1  185617SU8  185618SU6  186591SU4  187007SU0  187016SU1  187017SU9  187582SU2  187586SU3  187587SU1  187589SU7  188118SU4  188126SU7  188126SU7  188129SU1  188130SU9  189178SU7  189179SU5  189181SU1  189182SU9  189184SU5  189690SU1  189694SU3  189695SU0  189696SU8  189923SU6  190375SU6  190378SU0  190379SU8  190386SU3  190389SU7  190394SU7  190397SU0  190398SU8  190402SU8  190408SU5  191011SU6  191012SU4  191066SU0  191067SU8  191464SU7  191467SU0  191468SU8  191469SU6  191470SU4  191533SU9  191542SU0  191548SU7  191551SU1  191552SU9  191553SU7  191555SU2  191556SU0  191557SU8  191565SU1  191566SU9  191569SU3  191680SU8  191681SU6  192554SU4  192555SU1  192556SU9  192557SU7  192558SU5  192559SU3  192560SU1  192561SU9  192562SU7  192563SU5  192564SU3  192565SU0  192566SU8  192567SU6  192708SU6  192709SU4  192929SU8  192953SU8  193447SU0  193901SU6  193928SU9  193948SU7  193951SU1  193953SU7  193955SU2  193956SU0  193958SU6  193959SU4  193961SU0  193962SU8  194180SU6  194183SU0  194184SU8  194188SU9  194189SU7  194195SU4  194196SU2  194197SU0  194523SU7  194524SU5  194525SU2  194529SU4  194530SU2  194531SU0  194536SU9  194541SU9  194543SU5  195859SU4  195860SU2  195868SU5  195869SU3  195870SU1  195873SU5  196262SU0  196269SU5  196270SU3  196883SU3  196883SU3  197241SU3  197244SU7  197247SU0  197248SU8  197250SU4  197251SU2  197252SU0  197253SU8  197257SU9  197258SU7  197943SU4  198318SU8  198319SU6  198320SU4  198324SU6  198325SU3  198326SU1  198409SU5  198421SU0  198422SU8  198423SU6  198425SU1  198426SU9  198428SU5  198429SU3  198430SU1  198431SU9  198432SU7  198433SU5  198434SU3  198435SU0  198437SU6  198438SU4  198878SU1  198880SU7  198881SU5  198882SU3  198885SU6  198887SU2  198889SU8  198890SU6  198891SU4  198894SU8  198895SU5  198896SU3  198898SU9  198899SU7  198914SU4  198916SU9  198927SU6  198935SU9  198936SU7  198937SU5  198938SU3  199004SU3  199005SU0  199006SU8  199007SU6  199008SU4  199009SU2  199011SU8  199012SU6  199013SU4  199015SU9  199016SU7  199017SU5  199018SU3  199019SU1  199020SU9  199021SU7  199022SU5  199023SU3  199492SU0  199493SU8  199494SU6  199495SU3  199498SU7  199499SU5  199500SU0  199501SU8  199502SU6  199503SU4  199504SU2  199505SU9  199506SU7  199607SU3  199608SU1  199609SU9  199610SU7  199611SU5  199612SU3  199616SU4  199617SU2  199618SU0  199619SU8  199620SU6  200174SU1  200345SU7  201216SU9  201217SU7  201218SU5  201219SU3  201220SU1  201221SU9  201222SU7  201223SU5  201224SU3  201224SU3  201225SU0  201227SU6  201228SU4  201229SU2  201230SU0  201231SU8  201232SU6  201233SU4  201440SU5  201445SU4  201447SU0  201448SU8  201449SU6  201450SU4  201451SU2  201452SU0  201453SU8  201454SU6  201455SU3  201456SU1  201457SU9  201458SU7  201459SU5  201460SU3  201461SU1  201462SU9  201463SU7  201962SU8  201963SU6  201964SU4  201970SU1  202047SU7  202724SU1  203044SU3  203045SU0  203124SU3  203279SU5  203313SU2  203339SU7  203448SU6  203449SU4  203451SU0  203452SU8  203453SU6  203457SU7  203460SU1  203617SU6  203712SU5  203718SU2  203719SU0  203720SU8  203721SU6  203722SU4  203723SU2  203724SU0  203887SU5  203888SU3  203889SU1  203890SU9  203891SU7  203892SU5  203893SU3  203894SU1  203895SU8  203896SU6  203899SU0  203900SU6  203902SU2  203903SU0  203904SU8  203905SU5  203907SU1  203908SU9  203910SU5  203911SU3  203912SU1  203913SU9  203914SU7  203985SU7  203986SU5  204250SU5  204252SU1  204253SU9  204254SU7  204255SU4  204256SU2  204257SU0  204258SU8  204259SU6  204260SU4  204261SU2  204262SU0  204263SU8  204264SU6  204269SU5  204270SU3  204271SU1  204272SU9  204273SU7  204274SU5  204572SU2  204573SU0  204574SU8  204575SU5  204576SU3  204577SU1  204578SU9  204579SU7  204580SU5  204581SU3  204582SU1  204583SU9  204584SU7  204585SU4  204586SU2  204587SU0  204588SU8  204726SU4  204727SU2  204906SU2  204907SU0  204908SU8  204909SU6  204910SU4  204911SU2  204912SU0  204913SU8  204914SU6  204915SU3  204916SU1  204917SU9  204918SU7  204919SU5  204920SU3  204926SU0  204927SU8  205520SU0  205521SU8  205522SU6  205523SU4  205524SU2  205525SU9  205526SU7  205527SU5  205548SU1  205549SU9  205550SU7  205551SU5  205552SU3  205553SU1  205554SU9",
      "center_classification_date": "20130411",
      "distribution_pattern": "Worldwide Distribution, USA, including the states of AL, AZ, AR, CA, CO,K CT, DC, FL, GA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, MT  NE,  NV N NJ, NM, NY, NC, ND, OH, OK, PA, RI ,SC, SD, TN, TX, VA, WA, WI and the countries of AUSTRALIA, AUSTRIA, BRAZIL, CANDA, CHILE, CHINA, COSTA ARICA, CROATIA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY INDIA, INDONESIA ITALY JAPAN, KOREA, LATVIA, LEBANON, MALAYSIA, MEXICO, MOLDOVA, MOROCCO, NETHERLANDS, NEW ZEALAND, NORWAY, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA , SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM.",
      "state": "WI",
      "product_description": "GE Healthcare LOGIQ S8 (with software revisions R1.1.1 and R1.5.1) and LOGIQ S7 (with software revisions R1.0.1, R1.0.2 and R1.0.3) diagnostic ultrasound system with the S4-10-D Ultrasound Probe.  The device is intended for use in ultrasound evaluation.",
      "report_date": "20130417",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GE Healthcare, LLC",
      "recall_number": "Z-1096-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64725",
      "termination_date": "20131121",
      "more_code_info": "",
      "recall_initiation_date": "20130208",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}