{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Croissy Beaubourg",
      "state": "",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64589",
      "recalling_firm": "Trophy Sas",
      "address_1": "4 Rue Pelloutier",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "USA Nationwide Distribution",
      "recall_number": "Z-1094-2013",
      "product_description": "CareStream Dental Kodak 2100 Intraoral X-ray system    Product Usage:  The expected usage (intended use) for all of these products is for dental radiographic examination and diagnosis of the teeth, jaw and oral structures.",
      "product_quantity": "167 installed in the US",
      "reason_for_recall": "There have been approximately 2% of the x-ray generators manufactured between February 2012 and November 2012 returned due to errors. These generators showed the type of defect that indicates poor contact of HT transformer power pin. The defect is a bent connector associated with the generator.",
      "recall_initiation_date": "20120918",
      "center_classification_date": "20130502",
      "termination_date": "20170502",
      "report_date": "20130508",
      "code_info": "ABYJ011  ABYJ012  ABYJ020 through ABYJ026  ACJY522  ACYJ004 through ACYJ007  ACYJ009 through ACYJ012  ACYJ014  ACYJ015  ACYJ017 through ACYJ023  ACYJ500 through ACYJ509  ACYJ517 through ACYJ521  ACYJ523 through ACYJ531  ACYJ533 through ACYJ536  ADYJ001 through ADYJ005  ADYJ007  ADYJ008  ADYJ011 through ADYJ013  ADYJ015 through ADYJ018  ADYJ020 through ADYJ032  ADYJ034 through ADYJ037  ADYJ039 through ADYJ049  ADYJ051  ADYJ054  ADYJ056  ADYJ061  ADYJ062  ADYJ064 through ADYJ067  ADYJ069 through ADYJ073  ADYJ075 through ADYJ082  AEYJ001 through AEYJ003  AEYJ005  AEYJ006  AEYJ008 through AEYJ010  AEYP003 through AEYP011  AEYP024 through AEYP026  AEYP028 through AEYP030  AFYP005 through AFYP008  AFYP012  AFYP014 through AFYP018  AFYP021  AFYP022  AGYP010  AGYP012  AGYP016  AGYP018 through AGYP020  AGYP022  AGYP025  AGYP027  AGYP028  AGYP030  AGYP031  AGYP033 through AGYP041  AGYP045  AGYP046  AGYP068  AGYP069  AGYP073  AGYP086  AGYP088  AGYP089  AGYP092  AJYP025  AJYP027  AKYP001  AKYP007  AKYP009",
      "more_code_info": ""
    }
  ]
}