      "status": "Ongoing",
      "city": "Austin",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90023",
      "recalling_firm": "Luminex Corporation",
      "address_1": "12212 Technology Blvd",
      "address_2": "N/A",
      "postal_code": "78727-6101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US:OH, ME, PA, FL, IL, NE, IN, TX, GA, WA, MI, CO, AZ, OH, KY, CA, VT OUS: None",
      "recall_number": "Z-1092-2022",
      "product_description": "Verigene Processor SP, Models 10-0000-07/10-0000-07R  a component of the Verigene system",
      "product_quantity": "24 units",
      "reason_for_recall": "Heater  within the Verigene System may be outside of established temperature limits for assays which could result in false negative results.",
      "recall_initiation_date": "20220405",
      "center_classification_date": "20220513",
      "report_date": "20220525",
      "code_info": "Serial Numbers for 10-0000-07: 17340003 19044007 20133004 18128002 18214004 12061048 18191010 13193254 13051068 21029902 16277059 19233005 11250096 14069051 18255003   *18039010   *20212006  *19317004   *19206009   *18057008  Serial Numbers added 6.23.22 for 10-0000-07R: 18191010  13051068  14069051  *12198198   *12278344   *14322267  *12271338  Serial Numbers added 7.14.22 for 10-0000-07R: 18226007 and 16242043 Serial Numbers added 7.14.22 for 10-0000-07:  17062030, 18235008 and 19192006",
      "more_code_info": ""
    }
  ]
}