  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "345 E Main St",
      "reason_for_recall": "This single manufacturing lot has the potential for the bond strength between the Trabecular Metal material and substrate to be below the minimum required specification.  Testing found bond strength approximately 80% of the minimum required by specification in one of the three zones of the Trabecular Metal.  The other two zones exceeded the minimum specification.",
      "address_2": "",
      "product_quantity": "10 units",
      "code_info": "lot 62267250",
      "center_classification_date": "20140225",
      "distribution_pattern": "Distribution USA in the state of Illinois.",
      "state": "IN",
      "product_description": "00434811113,  REF I4348-111-13  Zimmer Trabecular Metal  Shoulder  Humeral Stem 48 Degrees  11 mm Stem Diameter 130 mm Stem Length Trabecular Metal Humeral Stem 48 Degree 11mm x 130mm,   Sterile, QTY 1, Rx    Product Usage:  Zimmer Trabecular Metal TM Humeral Stem is a shoulder prosthetic replacement device that is indicated for the treatment of severe pain or significant disability in degenerative, rheumatoid, or traumatic disease of the glenohumeral joint as well as humeral fractures. The humeral stem uses. Trabecular Metal around the proximal part of the stem for biological ingrowth.",
      "report_date": "20140305",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer, Inc.",
      "recall_number": "Z-1092-2014",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "67067",
      "termination_date": "20140911",
      "more_code_info": "",
      "recall_initiation_date": "20140110",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}