      "status": "Terminated",
      "city": "Tuusula",
      "state": "",
      "country": "Finland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67342",
      "recalling_firm": "Instrumentarium Dental, PaloDEx Group Oy",
      "address_1": "Nahkelantie 160",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) including states of: AK, AZ, CA, CO, GA, IL, KY, MD, MT, NC, NV, OR, PA, TN, TX, UT, VA, WA,and  WI; and Internationally to: Belgium Brazil, Chile, China, Czech Republic, France, Germany, Italy, Spain, Sweden, Switzerland, and United Kingdom.",
      "recall_number": "Z-1091-2014",
      "product_description": "The FOCUS is a microprocessor controlled intra-oral x-ray unit with a Load Matching Frequency (LMF) DC generator, which produces dental images on digital and film media. For the digital images the FOCUS is connected to the Sigma USB terminal which is connected to a PC, which reads the image data from a sensor and displays the image on the PC monitor. In order to make installation and service faster and easier the electronics is integrated into the horizontal arm.    FOCUS is intended to be used for producing diagnostic x-ray radiographs of dentition, jaws and other oral structures.",
      "product_quantity": "101 units (US Distribution)",
      "reason_for_recall": "Instrumentarium Dental has delivered a number of FOCUS 3.0 Intraoral X-ray units in November 2013 with a potential flaw in the mounting hardware. A number of wall mount units were delivered with FOCUS X-ray machines that had not been properly welded at our supplier. As a result of the incomplete welding operation, the wall mount unit on the FOCUS X-ray unit may not properly support the unit as expected, which could cause the unit to drift. Depending on the extent of the shortened or missing welds, there is also a possibility that the wall mount unit will fail to support the weight of the unit, resulting in the FOCUS X-ray unit falling unexpectedly from its mounted position.",
      "recall_initiation_date": "20140114",
      "center_classification_date": "20140225",
      "termination_date": "20150522",
      "report_date": "20140305",
      "code_info": "Device Model Number / Sales Code: 50600-IMG;     Serial Numbers (US and Foreign): F21221 - F21641, and F21643;     Serial Numbers (US Devices): F21221 - F21230, F21265 - F21270, F21273 - F21290, F21296 - F21350, F23154 - F21365, F21367 - F21391, F21395 - F21447, F21451 - F21464, F21469 - F21478, F21481 - F21486, F21490 - F21519, F21525 - F21542, F21545 - F21550, F21565 - F21588, F21623 - F21628, F21636 - F21641, and F21643",
      "more_code_info": ""
    }
  ]
}