  "results": [
    {
      "status": "Terminated",
      "city": "Broomfield",
      "state": "CO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70376",
      "recalling_firm": "Biomet Spine, LLC",
      "address_1": "310 Interlocken Pkwy Ste 120",
      "address_2": "N/A",
      "postal_code": "80021-3464",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed to TX, OH, NE and NY.",
      "recall_number": "Z-1089-2015",
      "product_description": "Telluride Percutaneous Rod Inserter Long, Model #7706-1904.  The Inserter is a reusable device provided within an aluminum sterilization case along with other orthopedic manual surgical instruments.  The case consists of three trays of instruments, and the middle tray contains a quantity of two Inserters.    The Inserter is for use as part of a Telluride System pedicle screw procedure.",
      "product_quantity": "10",
      "reason_for_recall": "The Telluride Percutaneous Rod Inserter Long does not contain a feature that locks the Rod in place during use.  There is a potential for the Rod to disengage from the Inserter during insertion, and could lead to harm to the patient.",
      "recall_initiation_date": "20111219",
      "center_classification_date": "20150210",
      "termination_date": "20150323",
      "report_date": "20150218",
      "code_info": "LOT Numbers: L561968 and L570545  Product Code: 7703-1600"
    }
  ]
}