  "results": [
    {
      "status": "Terminated",
      "city": "Littleton",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73303",
      "recalling_firm": "Medtronic Navigation, Inc.",
      "address_1": "300 Foster St",
      "address_2": "N/A",
      "postal_code": "01460-2017",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution to Austria only.",
      "recall_number": "Z-1088-2016",
      "product_description": "Medtronic 0-arm 02 Imaging System  Catalog Number: Bl-700-02000  The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects.    The O-arm O2 Imaging System is compatible with certain image guided surgery systems.",
      "product_quantity": "1 unit",
      "reason_for_recall": "Screws holding the detector panel in as a result  the detector panel could make contact with other parts with in the 0-arm¿ 02 Imaging System, causing noise and vibration when being positioned or during 3D Imaging could lead to poor image quality",
      "recall_initiation_date": "20160129",
      "center_classification_date": "20160308",
      "termination_date": "20160310",
      "report_date": "20160316",
      "code_info": "Serial Number: C0936"
    }
  ]
}