  "results": [
    {
      "status": "Terminated",
      "city": "Algete",
      "state": "N/A",
      "country": "Spain",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73428",
      "recalling_firm": "Sedecal S.A.",
      "address_1": "Pol. Industrial Rio De Janiero",
      "address_2": "Calle Pelaya 9-13",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NM, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI.",
      "recall_number": "Z-1084-2016",
      "product_description": "Mobile wDR motorized portable diagnostic X-ray systems.  Intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts on both adult and pediatric patients. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not intended for mammography.",
      "product_quantity": "349 systems",
      "reason_for_recall": "This is related to previous 806 Notifications about uncommanded movements of a motorized mobile x-ray system. Reason for Recall: Along with the previously noted encoder and gauges problems and corresponding fixes, we have observed that static electricity can also elicit the unwanted behavior.",
      "recall_initiation_date": "20160118",
      "center_classification_date": "20160308",
      "termination_date": "20180207",
      "report_date": "20160316",
      "code_info": "Serial Numbers: 485886/SN11000007, 485062/SN11000009, 488169/SN11000011, 485061/SN11000012, 485059/SN11000013, 485060/SN11000015, 489089/SN11000016, 489616/SN11000033, 490027/SN11000034, 489694/SN11000035, 489695/SN11000036, 489696/SN11000037, 490221/SN11000039, 489088/SN11000040, 493566/SN11000047, 493565/SN11000048, 494867/SN11000051, 494941/SN11000052, 492417/SN11000053, 494456/SN11000055, 494865/SN11000056, 494864/SN11000060, 495286/SN11000061, 494866/SN11000062, 494464/SN11000065, 497570/SN11000066, 488171/SN11000068, 494060/SN11000070, 494061/SN11000071, 494062/SN11000072, 494063/SN11000073, 494064/SN11000074, 494065/SN11000075, 497715/SN11000080, 496446/SN11000082, 496387/SN12000003, 495938/SN12000007, 495939/SN12000008, 493364/SN12000022, 493462/SN12000023, 501028/SN12000026, 500739/SN12000027, 500740/SN12000028, 502424/SN12000034, 503143/SN12000036, 503142/SN12000037, 503141/SN12000038, 503144/SN12000039, 503541/SN12000040, 500466/SN12000045, 502721/SN12000047, 505077/SN12000048, 506626/SN12000051, 507607/SN12000054, 507906/SN12000055, 508342/SN12000057, 500468/SN12000058, 496401/SN12000059, 500467/SN12000060, 508838/SN12000061, 508532/SN12000062, 504450/SN12000064, 509718/SN12000068, 510607/SN12000072, 512160/SN12000077, 513477/SN12000080, 513478/SN12000081, 512892/SN12000082, 513848/SN12000084, 512887/SN12000085, 512888/SN12000086, 512890/SN12000087, 508699/SN12000088, 514462/SN12000089, 514461/SN12000090, 514104/SN12000091, 514103/SN12000092, 509062/SN12000107, 508783/SN12000108, 516891/SN12000110, 516896/SN12000111, 497572/SN12000112, 518135/SN12000118, 518342/SN12000119, 518343/SN12000120, 518344/SN12000121, 518345/SN12000122, 518346/SN12000123, 502025/SN12000124, 502026/SN12000125, 12000166, 519326/SN13000005, 519388/SN13000006, 519511/SN13000010, 519506/SN13000011, 519387/SN13000012, 519763/SN13000013, 519434/SN13000014, 520251/SN13000016, 520253/SN13000017, 519762/SN13000018, 520254/SN13000019, 520255/SN13000020, 520256/SN13000021, 520413/SN13000023, 520414/SN13000025, 520415/SN13000026, 520252/SN13000027, 520422/SN13000028, 520418/SN13000030, 520416/SN13000031, 520417/SN13000033, 522633/SN13000035, 522944/SN13000037, 523279/SN13000042, 523349/SN13000043, 13000046, 13000047, 13000048, 13000049, 13000053, 13000054, 13000055, 13000056, 13000057, 13000058, SN13000059, 13000060, SN13000061, SN13000062, SN13000063, 13000067, SN13000069, 13000070, 13000071, SN13000072, SN13000073, SN13000074, 13000075, 13000076, 13000077, 13000078, 13000079, SN13000080, SN13000082, SN13000083, SN13000084, 13000085, SN13000087, SN13000089, SN13000090, SN13000091, 13000092, 13000093, 13000094, SN13000099, SN13000100, SN13000101, SN13000102, SN13000103, SN13000104, SN13000106, SN13000107, 13000109, SN13000112, SN13000113, 519598/SN13000118, 519599/SN13000119, 519600/SN13000120, SN13000125, 13000127, 13000139, 13000140, 13000141, 13000143, 13000144, 13000145,  SN13000146, SN13000147, SN13000148, SN13000149, SN13000153, SN13000154, SN13000157,  SN7369686, SN13000159, SN13000160, SN13000163, SN13000164, SN13000165, 13000167, 13000168, 13000169, SN13000170,  SN13000174,       SN13000176,  SN13000177,  SN13000179,  SN13000180,   SN13000181, SN13000182,  SN13000183,  SN13000185, 13000196, 13000199, 13000200, 13000201, 13000202, 13000203, 13000204, 13000205,  SN13000206, 14000004, 14000005, 14000006, 14000008, 14000009, 14000010, 14000011, 14000013, 14000014, 14000015, 14000016, 14000017, 14000019, 14000022, 14000024, 14000026, 14000028, 14000029, 14000031, 14000032, 14000033, 14000034, 14000037, 14000038, 14000041, 14000043, 14000046, 14000047, 14000048, 14000049, 14000050, 14000051, 14000052, SN14000061, 14000069, SN14000072, 14000075, 14000077, 14000078, 14000079, 14000080, 14000081, 14000082, 14000084, 14000086, 14000087, 14000088, 14000089, SN14000096, SN14000097, SN14000098, SN14000099, SN14000100, 14000092, 14000083, 14000085, 14000103,   14000106,   14000083,  14000109, 14000112, 14000116, 14000119, 14000118, 14000120, 14000121, 14000122, 14000124, SN14000128, 14000127, 14000126,  SN14000130, 14000134, 14000135, 14000138, 14000139, SN15000002, 14100044, 14100045, 14100053, 14100054, 14100085, 14100086, 14100087, 14100088, 14100090, 14100094, 14100095, 14100097, 14100101, 14100104, 14100105, 14100109, 14100111, 14100117, 14100126, 14100127, 14100128,  SN14100046, 15100019, 15100023, SN14100047, SN14100055, SN14100071, SN14100072, SN14100073, SN14100074, SN14100075, 15100028, 15100034, 15100035, 15100036, 15100039, 15100040, 15100041, 15100042, 15100047, 15100048, 15100050, 15100071, 15100084, 15100087, 15100088, 15100089, 15100091, 15100092, 15100093, 15100094, 15100095, 15100097, 15100098, 15100101, 15100103, 15100104,   SN15100020,  SN15100009,  SN15100016, SN15100010, SN15100011, SN15100013 and SN15100018."
    }
  ]
}