  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Glendora",
      "address_1": "1332 S Lone Hill Ave",
      "reason_for_recall": "Ormco Corporation is voluntarily recalling one (1) lot of Bracket Buccal Tubes as a result of a packaging error. More specifically, the Bracket  Buccal Tubes product package correctly identifies the product as gingival (0¿) offset brackets, but  actually contains distal (10¿) offset brackets. Use of the affected product may cause unintended rotation  of the molars. Although reversible, a delay in",
      "address_2": "",
      "product_quantity": "8210",
      "code_info": "Part Number    438-0184             Lot Number  051063597",
      "center_classification_date": "20130410",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including the country of EU.",
      "state": "CA",
      "product_description": "Bracket Buccal Tube U7L DB/UL22 Roth MT HK 0/-14/0 \t      Product Usage:  Bracket Buccal Tubes are intended to hold in orthodontic wires used to move teeth during orthodontic treatment",
      "report_date": "20130417",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Ormco/Sybronendo",
      "recall_number": "Z-1084-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64749",
      "termination_date": "20130904",
      "more_code_info": "",
      "recall_initiation_date": "20130308",
      "postal_code": "91740-5339",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}