  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Harleysville",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70351",
      "recalling_firm": "Medical Components, Inc dba MedComp",
      "address_1": "1499 Delp Dr",
      "address_2": "N/A",
      "postal_code": "19438",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-1083-2015",
      "product_description": "Hemodialysis Catheters and PICCs: Product identification: MCDLTSL24P  (SF X 24CM LT HEMO-CATH), MC061422 (14FX15CM SLX DBL LMN ST CATH),   MRCTP41017 (4F X 55CM SINGLE CT PICC), MRCTP52007 (5F DUAL PRO-PICC NURSING TRAY), MRCTP52010 (5F DUAL PRO-PICC MAX BARR TRAY),   MR81013106MB  (3FX60CM SGL TAPRLS VASCU-PICC).    Intended uses: To attain long-term vascular access for hemodialysis  and apheresis.     The Medcomp   SLX Silicone Double Lumen Catheter is designed for acute hemodialysis  and apheresis. The Pro-PICC¿ CT catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media, and allows for central venous pressure monitoring. The Peripherally Inserted Central Vein Access Catheters are designed for Long or Short- Term peripheral access to the central venous system for intravenous therapy and blood sampling, and allows for central venous pressure monitoring.",
      "product_quantity": "243",
      "reason_for_recall": "Drug products contained within the kits may have been rendered ineffective by a duplicate sterilization process. The kits include one or more  of the  following drugs: 5ml  ampule  of  lidocaine,  3ml  Chloraprep  one-step, 10ml  pre-filled saline syringe, Hygea saline wipes, and 3ml tinted Chloraprep. No injuries or illnesses were reported.",
      "recall_initiation_date": "20150121",
      "center_classification_date": "20150210",
      "termination_date": "20150803",
      "report_date": "20150218",
      "code_info": "Lot numbers: MBWK970, MBWH890, MBWJ180, MBWJ460, MBWH910, MBWH940"
    }
  ]
}