  "results": [
    {
      "country": "United States",
      "city": "Waukesha",
      "address_1": "3000 N Grandview Blvd",
      "reason_for_recall": "GE Healthcare has become aware of an issue associated with the base casting for the articulated arm of your VMX and  VMX Plus Mobile X-ray systems that may impact Operator or Patient safety.",
      "address_2": "",
      "product_quantity": "1296",
      "code_info": "FMI 10850: VMX ARM BASE CASTING REPLACEMENT  Serial Number  00000155075MO5  00000006136YY6  00000006802YY3  00000006799YY1  00000007989YY7  00000007990YY5  00000002358YY0  00000021131YY8  00000021132YY6  00000021146YY6  00000021139YY1  00000021141YY7  00000021159YY9  00000041750YY1  00000021108YY6  00000041725YY3  00000041778YY2  00000083296YY4  00000085032YY1  00000118252MO6  00000085956YY1  00000160478MO4  00000021116YY9  00000021017YY9  00000011088YY2  00000024550YY6  00000021097YY1  00000031512YY7  00000031585YY3  00000025793YY1  00000002355YY6  00000006148YY1  00000147625MO8  00000021024YY5  00000021134YY2  00000021140YY9  00000024543YY1  00000021103YY7  00000021107YY8  00000021151YY6  00000041762YY6  00000041772YY5  00000041768YY3  00000041766YY7  00000041764YY2  00000041771YY7  00000041770YY9  00000041784YY0  00000041783YY2  00000083295YY6  00000140455MO7  00000140451MO6  00000086024YY7  00000086043YY7  00000086036YY1  00000086025YY4  00000086046YY0  00000086048YY6  00000140453MO2  00000086049YY4  00000086032YY0  00000086026YY2  00000086033YY8  00000140467MO2  00000140473MO0  00000140459MO9  00000140464MO9  00000140460MO7  00000140471MO4  00000160470MO1  00000008077YY0  00000083292YY3  00000059632YY1  00000006796YY7  00000006804YY9  00000021157YY3  00000021158YY1  00000041780YY8  00000086001YY5  00000086016YY3  00000086017YY1  00000140487MO0  00000021051YY8  00000155065MO6  00000085037YY0  00000085031YY3  00000084999YY2  00000106771MO9  00059632YY70CS  00000140535MO6  00059632YY71CS  00000005327YY2  00000013802YY4  00000024542YY3  00000024564YY7  00000031588YY7  00000048207YY5  00000048221YY6  00000048220YY8  00000059601YY5  00000083261YY8  00000085043YY8  00000085060YY2  00000085081YY8  00000085080YY0  00000089505MO2  00106782MO35CS  00000089509MO4  00000089520MO1  00000089522MO7  00000089532MO6  00106782MO37CS  00000089547MO4  00000106753MO7  00000118276MO5  00000118277MO3  00000118279MO9  00000118278MO1  00000140525MO7  00059632YY72CS  00000140526MO5  00000140546MO3  00000147624MO1  00000106778MO4  00000106780MO0  00000155024MO3  00000041752YY7  00000025765YY9  00000059632YY3  00000048275YY2  00000106782MO8  00000106769MO3  00000155057MO3  00000010905YY8  00000011006YY4  00000013795YY0  00000140503MO4  00000140510MO9  00000147655MO5  00000155022MO7  00000008195YY0  00000011112YY0  00000021000YY5  00000008116YY6  00000008117YY4  00000011046YY0  00000023060YY7  00000086021YY3  00000005259YY7  00000006765YY2  00000022983YY1  00000022996YY3  00000010898YY5  00000002303YY6  00000005295YY1  00000005260YY5  00000011047YY8  00000002322YY6  00000010942YY1  00000023020YY1  00106782MO54CS  00000010952YY0  00000022997YY1  00000024574YY6  00000013817YY2  00000011017YY1  0059632YY123CS  00000085020YY6  00000005239YY9  00000005267YY0  0059632YY103CS  00059632YY94CS  00000085030YY5  00000011019YY7  00000008003YY6  00000005243YY1  00000022982YY3  00000014353YY7  00000021098YY9  00000007963YY2  00000023061YY5  00000007957YY4  00000007956YY6  00000023037YY5  0059632YY105CS  00000021063YY3  00000021092YY2  00000021091YY4  00000002349YY9  00000002330YY9  00000002329YY1  00000008133YY1  00000008132YY3  00000021088YY0  00000002317YY6  00000002318YY4  00000006791YY8  00000006793YY4  00000011070YY0  00000006779YY3  00000008141YY4  00000002340YY8  00000002282YY2  00000008196YY8  00000005250YY6  00000005352YY0  00000008200YY8  00000008199YY2  00000008001YY0  00000013816YY4  00000086022YY1  00000005287YY8  00000036034YY6  00000005324YY9  00000005253YY0  00000005254YY8  00000011102YY1  00000001904YY2  00000083284YY0  00000002308YY5  00000024553YY0  00000010956YY1  00000024577YY9  00000024582YY9  00000013805YY7  00000010993YY4  00000024555YY5  00000011122YY9  00000008185YY1  00000008184YY4  00000085958YY7  00000002288YY9  00000008013YY5  00000013814YY9  00000013808YY1  00000021023YY7  00000005315YY7  00000022985YY6  00000023039YY1  00000021052YY6  00000023055YY7  00000005310YY8  00000005309YY0  00000010979YY3  00000021012YY0  00000021013YY8  00106782MO64CS  00000011092YY4  00000005353YY8  00000011103YY9  00000014338YY8  00000005240YY7  00000006787YY6  00000014344YY6  00000011045YY2  00000002306YY9  00000023029YY2  00000001913YY3  00000023005YY2  00000007967YY3  00000083293YY1  00000106837MO8  00000007970YY7  00000002314YY3  00000014347YY9  00000085137YY8  00000022986YY4  00106782MO58CS  00000022995YY5  00106782MO59CS  00000024586YY0  00000025730YY3  00000024592YY8  00000024594YY4  00000024604YY1  00000025770YY9  00000024615YY7  00000025775YY8  00000025781YY6  00000031500YY2  00000025760YY0  00000031536YY6  00000025789YY9  00000025768YY3  00000031514YY3  00000106777MO6  00000086039YY5  00000086045YY2  00000025803YY8  00000025805YY3  00000025804YY6  00000086010YY6  00000083281YY6  00000031516YY8  00000085958YY8  0059632YY121CS  00000031545YY7  00000041704YY8  00000031543YY2  00000031544YY0  00000085024YY8  00000031552YY3  00000031537YY4  0059632YY112CS  00000031570YY5  0059632YY111CS  00000031575YY4  00000041712YY1  00000031572YY1  00000031569YY7  00000031576YY2  00000031583YY8  00000041723YY8  00000048200YY0  00000048222YY4  00000041744YY4  00000041741YY0  00000048223YY2  00000041745YY1  00000041747YY7  00000041748YY5  00000041746YY9  00000041742YY8  00000048252YY1  00000048258YY8  00000048257YY0  0059632YY102CS  00000048238YY0  00000048260YY4  00000048264YY6  00000048236YY4  00000048242YY2  00000048290YY1  0059632YY101CS  00000048274YY5  00000048292YY7  00000048296YY8  00000059957YY1  00000059964YY7  00000008026YY7  00000059612YY2  00000085119YY6  00000059965YY4  00000059967YY0  00000059618YY9  00106782MO51CS  00000002339YY0  00000011123YY7  00000011079YY1  0059632YY122CS  00106782MO61CS  00000021078YY1  00000008111YY7  00000106811MO3  00000006758YY7  00000059971YY2  00000002293YY9  00000059977YY9  00000086029YY6  00000086030YY4  00000083272YY5  00000083260YY0  00000083270YY9  00000059987YY8  00000011126YY0  0059632YY114CS  0059632YY100CS  00000083276YY6  00000008181YY0  00000085095YY8  00000085007YY3  00000085006YY5  00000014319YY8  00000013821YY4  00000085050YY3  00000085107YY1  0059632YY118CS  00106782MO72CS  00106782MO73CS  0059632YY115CS  0059632YY125CS  00000085058YY6  00000085109YY7  00000085112YY1  00000085116YY2  00000023057YY3  00000085118YY8  00000085072YY7  00000085075YY0  00000085074YY3  00000085136YY0  00000085141YY0  00059632YY98CS  00000085147YY7  00000089569MO8  00059632YY96CS  00000089515MO1  00000089508MO6  00000089598MO7  00000089516MO9  00000089519MO3  00000011091YY6  00000089602MO7  00000085065YY1  00000106773MO5  00000089571MO4  0059632YY110CS  00000089575MO5  00000089523MO5  00000089518MO5  00000106801MO4  00000089527MO6  00000106802MO2  00000106812MO1  00059632YY93CS  00000106820MO4  00000106825MO3  00000106836MO0  00000106840MO2  00000106827MO9  00000106828MO7  0059632YY108CS  00059632YY95CS  00000106835MO2  00000106833MO7  00000106834MO5  00106782MO74CS  00000106751MO1  00000085019YY8  00000118283MO1  00000118271MO6  00000118287MO2  00000140511MO7  00000083297YY2  00000008109YY1  00000140528MO1  00000140529MO9  00059632YY99CS  00106782MO56CS  00000147664MO7  00000147644MO9  00000155023MO5  00000155025MO0  00000155074MO8  00000008086YY1  00000155055MO7  00000085036YY2  0059632YY113CS  0059632YY104CS  00000008018YY4  0059632YY117CS  0059632YY116CS  00000023047YY4  00000085105YY5  00000089590MO4  00000007937YY6  00000025799YY8  0059632YY107CS  00000010987YY6  00000041729YY5  0000106782MO70  00106782MO75CS  00000025769YY1  00000089573MO0  00000155027MO6  00000022976YY5  00000024525YY8  00000031513YY5  00106782MO65CS  00059632YY97CS  00000106768MO5  0059632YY109CS  00000021065YY8  00000011128YY6  00000011081YY7  00059632YY80CS  00000007952YY5  00000023052YY4  00000006774YY4  00000006775YY1  00000006771YY0  00106782MO43CS  00000005340YY5  00000008058YY0  00000008015YY0  00000023031YY8  00059632YY81CS  00000010930YY6  00000010931YY4  00000010929YY8  00059632YY77CS  00000021070YY8  00000010971YY0  00000008187YY7  00059632YY78CS  00000022994YY8  00000021081YY5  00000023059YY9  00000021083YY1  00000021068YY2  00000008060YY6  00059632YY79CS  00000023018YY5  00000011104YY7  00000024565YY4  00000011030YY4  00000008016YY8  00000022993YY0  00059632YY84CS  00000005317YY3  00000022992YY2  00000006767YY8  00000006781YY9  00000006768YY6  00000006764YY5  00000006766YY0  00000024584YY5  00000025728YY7  00000025739YY4  00059632YY87CS  00000041733YY7  00000041732YY9  00000048224YY0  00000059982YY9  00059632YY89CS  00000083277YY4  00059632YY88CS  00059632YY86CS  00000085132YY9  00059632YY85CS  00000059632YY0  00059632YY83CS  00000140544MO8  00059632YY90CS  00000140531MO5  00000008017YY6  00000023034YY2  00000021075YY7  00000048269YY5  00000059954YY8  00000089584MO7  00000089601MO9  00000147635MO7  00000147627MO4  00000147639MO9  00000147643MO1  00000147648MO0  00000147651MO4  00000147670MO4  00000059632YY5  00000059632YY6  00000059632YY7  00000059632YY8  00000008129YY9  00000021010YY4  00000008004YY4  00000022975YY7  00000022973YY2  00000023001YY1  00000024595YY1  00000041728YY7  00000024521YY7  00000025811YY1  00000059974YY6  00000085017YY2  00000085093YY3  00000089567MO2  00000085122YY0  00000085104YY8  00059632YY73CS  00000089538MO3  00000140548MO9  00000140549MO7  00000008120YY8  00000106818MO8  00000002338YY2  00000011097YY3  00000011050YY2  00000023000YY3  00000002345YY7  00000005335YY5  00000005336YY3  00000031574YY7  00000085076YY8  00000085078YY4  00106782MO38CS  00106782MO39CS  00059632YY74CS  00000106782MO4  00106782MO41CS  00000106762MO8  00000059969YY6  00000059968YY8  00000025792YY3  00000083288YY1  00000085021YY4  00000140497M09  00000085955YY3  00000085954YY6  00000085022YY2  00000085023YY0  00000085027YY1  00000085957YY9  00000085924YY8  00000085953YY8  00000085952YY0  00000085024YY5  00000085028YY9  00000085951YY2  00000085026YY3  00000085059YY4  00000140524MO0  0059632YY135CS  00000011003YY1  00000014350YY3  00000023008YY6  00000118256MO7  00000147659MO7  00000005307YY4  00106782MO85CS  00000010998YY3  00000008128YY1  00000011039YY5  00000024547YY2  00000024545YY6  00000024546YY4  00000005286YY0  00000008172YY9  00000011037YY9  00000024528YY2  00000024569YY6  00000025720YY4  00000008182YY8  00000021028YY6  00000021029YY4  00000021030YY2  00000021031YY0  00000023009YY4  00106782MO86CS  00000011020YY5  00000010900YY9  00106782MO90CS  00000005271YY2  00000010939YY7  00000023063YY1  00000008179YY4  00000002347YY3  00000024520YY9  00000005350YY4  00000010962YY9  00000013791YY9  00000013790YY1  00000008124YY0  00106782MO92CS  00000022990YY6  0106782MO101CS  00000010955YY3  00000005245YY6  00000023032YY6  00106782MO96CS  00000005244YY9  00000021059YY1  00000005312YY4  00000155043MO3  00000024532YY4  00000005280YY3  00000005334YY8  00000011009YY8  00000013782YY8  00000011016YY3  00000013797YY6  00000005293YY6  00000013787YY7  00000013789YY3  00000013788YY5  00106782MO91CS  00000021053YY4  00000023011YY0  00000008131YY5  00106782MO83CS  00106782MO87CS  00000021040YY1  00000007964YY0  00000021005YY4  00000021021YY1  00000021006YY2  00000010995YY9  00106782MO97CS  00000024531YY6  00000023026YY8  00000005351YY2  00000005347YY0  00000021033YY6  00000008173YY7  00000005270YY4  00106782MO89CS  00000011053YY6  00000021018YY7  00000010906YY6  00000025745YY1  00000025743YY6  00000025744YY4  0059632YY139CS  00000025755YY0  00000025779YY0  00000025780YY8  00000025778YY2  00000025762YY6  00000025788YY1  00000025794YY9  00000025795YY6  00000025790YY7  00000031502YY8  00000031507YY7  00000025814YY5  00000031534YY1  00106782MO84CS  00000031509YY3  0059632YY142CS  00000041708YY9  00000031564YY8  00000031565YY5  00000031571YY3  00106782MO95CS  00000041718YY8  00000041719YY6  00000031587YY9  00000031578YY8  00000155071MO4  00000155073MO0  00000031599YY4  00000031598YY6  00000048201YY8  0059632YY138CS  00000041734YY5  00000048217YY4  00000041739YY4  00000048227YY3  00000048279YY4  00000048276YY0  00000048277YY8  00000048278YY6  00000048239YY8  00000048240YY6  00000048247YY1  00000048248YY9  00000048249YY7  00000048297YY6  00000059607YY2  00000059606YY4  00000059608YY0  00000059960YY5  00000059959YY7  00000048298YY4  00106782MO94CS  00000059958YY9  00000155060MO7  00000059613YY0  00000059627YY0  00000059622YY1  0059632YY136CS  00000059983YY7  00000059985YY2  00000059988YY6  00000083251YY9  00000059630YY4  00000083250YY1  00000085014YY9  00000085013YY1  00000085002YY4  00000085000YY8  00000085001YY6  00000085003YY2  00000085099YY0  00000085092YY5  00000085098YY2  00000085096YY6  00000085008YY1  00000085041YY2  00000085005YY7  00000085004YY0  00000085094YY1  00000085103YY0  00000085051YY1  00000085052YY9  00000085048YY7  00000085047YY9  00000085057YY8  00000085056YY0  00000085067YY7  00000085066YY9  00000085068YY5  00000085121YY2  00000085125YY3  00000085127YY9  00000085124YY6  00000085120YY4  00000085100YY6  00000085073YY5  00000085128YY7  00000059624YY7  00000085138YY6  00000085139YY4  00000085084YY2  00000085089YY1  00000089563MO1  00000089562MO3  00000089561MO5  00000089514MO4  00000085088YY3  00000089568MO0  00000089526MO8  00000089525MO0  00000106807MO1  00000089539MO1  00000106823MO8  00000106838MO6  00000106845MO1  00000118245MO0  00000118255MO9  00000118285MO6  00000118262MO5  00000106756MO0  00000106755MO2  00000106757MO8  00000106760MO2  00000106758MO6  00000106759MO4  00000118282MO3  00000118281MO5  00000118274MO0  00000118275MO7  00000118280MO7  00000155056MO5  00000147654MO8  00000155066MO4  00000025750YY1  0059632YY141CS  00106782MO99CS  00000155069MO8  00000085097YY4  00000025751YY9  0059632YY137CS  0059632YY140CS  00000155021MO9  00000155028MO4  00000155030MO0  00000155032MO6  00000155033MO4  00000155034MO2  00000155035MO9  00000155036MO7  00000048286YY9  00000048287YY7  00000048289YY3  00000048288YY5  00000021054YY2  00000085143YY6  00000085142YY8  00000085145YY1  00000085144YY4  00059632YY26CS  00059632YY31CS  00059632YY32CS  00059632YY27CS  00059632YY28CS  00059632YY33CS  00059632YY29CS  00059632YY30CS  00000024522YY5  00000005283YY7  00106782MO11CS  00000007997YY0  00000008006YY9  00000011111YY2  00000005341YY3  00000024602YY5  00000041705YY5  00000031533YY3  00000048245YY5  00000048250YY5  00000085062YY8  00000089537MO5  00000059632YY9  00059632YY10CS  00000023015YY1  00000106804MO8  00000106805MO5  00000118273MO2  00000008190YY1  00000011094YY0  00000048262YY0  00000048295YY0  00000006800YY7  00000008106YY7  00000021102YY9  00000021153YY2  00000086040YY3  00000024552YY2  00000106782MO9  00000041713YY9  00000085011YY5  14350  00000089570MO6  00000106806MO3  00000106809MO7  00059632YY12CS  00000118286MO4  00000140536MO4  00059632YY11CS  00000155026MO8  00059632YY13CS  00000024551YY4  00000011038YY7  00000005257YY1  00000106810MO5  00000024571YY2  0059632YY144CS  00000031528YY3  00000008110YY9  00000006151YY5  00000021133YY4  00000025773YY3  00000008162YY0  00000008051YY5  00000025800YY4  00000155070MO6  00000008119YY0  00000023016YY9  00000023017YY7  00000024567YY0  00000031555YY6  00000048241YY4  00000083255YY0  00000085102YY2  00000089557MO3  00000089510MO2  00000089513MO6  00000089574MO8  00000089550MO8  00059632YY75CS  00000106752MO9  00000010945YY4  00000010927YY2  00000013792YY7  00106782MO76CS  00000024611YY6  00000024612YY4  00000031511YY9  00000048232YY3  00000140502MO6  00000005248YY0  00000024599YY3  00000024598YY5  00000011076YY7  0059632YY128CS  0059632YY126CS  0059632YY127CS  00000025785YY7  00000085010YY7  00000085070YY1  00000106779MO2  00000147665MO4  00000022970YY8  00000021032YY8  00059632YY57CS  00000022989YY8  00000021036YY9  00000023058YY1  00106782MO21CS  00000031510YY1  00000025807YY9  00000031529YY1  00000031542YY4  00000031541YY6  00000031580YY4  00000048203YY4  00000031568YY9  00000048226YY5  00000083252YY7  00106782MO22CS  00106782MO23CS  00000085086YY7  00059632YY46CS  00059632YY47CS  00000089581MO3  00059632YY51CS  00059632YY56CS  00000089544MO1  00000106750MO3  00000118249MO2  00000118250MO0  00000089553MO2  00000118261MO7  00000118264MO1  00000118258MO3  00000140520MO8  00106782MO26CS  00000106763MO6  00000106765MO1  00000106764MO4  00000140550MO5  00000147633MO2  00000147631MO6  00106782MO28CS  00059632YY53CS  00059632YY54CS  00000106782MO2  00000147668MO8  00106782MO29CS  00000106775MO0  00000155039MO1  00000106781MO8  00000147673MO8  00000155014MO4  00000155079MO7  00000155059MO9  00106782MO19CS  00000155080MO5  00000155076MO3  00000106782MO6  00000155078MO9  00059632YY60CS  00000155041MO7  00000155045MO8  00000155042MO5  00000155044MO1  00059632YY58CS  00000106774MO3  00000118265MO8  00000118266MO6  00000118269MO0  00000118270MO8  00000140475MO5  00000118291MO4  00000140477MO1  00000140480MO5  00000140481MO3  00059632YY40CS  00000140493MO8  00000023054YY0  00059632YY39CS  00000140492MO0  00000140483MO9  00059632YY41CS  00000160465MO1  00000160466MO9  00000025741YY0  00000118288MO0  00000023041YY7  00000024558YY9  00000059617YY1  00000155058MO1  00000140507MO5  00000022981YY5  00000024535YY7  00000024559YY7  00000011048YY6  00000011049YY4  00000021071YY6  00000013807YY3  00000048206YY7  00000089521MO9  00000089528MO4  00106782MO42CS  00000140522MO4  00059632YY76CS  00000089596MO1  00000106826MO1  00000106846MO9  00000106848MO5  00000147662MO1  00000021039YY3  14345  00000025738YY6  00000089585MO4  00000089586MO2  00000089597MO9  00000106803MO0  00059632YY14CS  00059632YY20CS  00059632YY15CS  00059632YY21CS  00000118253MO4  00000118247MO6  00000118246MO8  00000118254MO2  00059632YY23CS  00059632YY16CS  00000106782MO1  00000118259MO1  00000085115YY4  00000021002YY1  00000041222BU1  00000010938YY9  00000002319YY2  00000011099YY9  00000005331YY4  00000024523YY3  00000084950EC5  00000007941YY8  00000005290YY2  00000021043YY5  00000002344YY0  00000023044YY1  00000014322YY2  00000011027YY0  00000011028YY8  00000022977YY3  0059632YY130CS  00000021027YY8  00000005332YY2  00000010974YY4  00000005261YY3  00000021090YY6  00000013806YY5  00000010928YY0  00000014323YY0  00000011095YY7  00000022974YY0  00000007998YY8  00000011056YY9  00000024533YY2  00000022984YY9  00000007999YY6  0059632YY132CS  00000023045YY8  00000023062YY3  00000024560YY5  00000025725YY3  00000024573YY8  00000024556YY3  00000024549YY8  00000024534YY0  00000025724YY6  00000025722YY0  00000022998YY9  00000024601YY7  00000024603YY3  00000025727YY9  00000025763YY4  00000025761YY8  00000025764YY2  00000025766YY7  00000031554YY9  00000031523YY4  00000031508YY5  00000031520YY0  00000031521YY8  00000031525YY9  00000031547YY3  00000031548YY1  00000031549YY9  00000031557YY2  00000031526YY7  00000031540YY8  00000031556YY4  00000025757YY6  00000031589YY5  00000031590YY3  00000031586YY1  00000059603YY1  00000048225YY7  00000048233YY1  00000048230YY7  00000048231YY5  00000048256YY2  00000025816YY0  00000048281YY0  00000048283YY6  00000048272YY9  00000048270YY3  00000048246YY3  00000048293YY5  00000059951YY4  00000059961YY3  00000048282YY8  00000059621YY3  00000140497MO9  00000059970YY4  00000059620YY5  00000083263YY4  00000085111YY3  00000083267YY5  00000085046YY1  00000140470MO6  00000140476MO3  00000085113YY9  00000059629YY6  00000085069YY3  00000085131YY1  00000059628YY8  00000089559MO9  00000085085YY9  00000089517MO7  00000089591MO2  00000089531MO8  00000089534MO2  00000106814MO7  0059632YY134CS  00000106824MO6  00000140458MO1  00000106841MO0  00000118244MO3  0059632YY129CS  00000118272MO4  00000089554MO0  00000140530MO7  00000140506MO7  00106782MO80CS  00000140469MO8  00000155015MO1  00000155016MO9  00000147634MO0  00000155017M07  00000014346YY1  00000155068MO0  00000024606YY6  00000140463MO1  00106782MO82CS  00000007959YY0  00000089524MO3  00000059962YY1  00000007966YY5  00000014354YY5  00000089564MO9  00000059973YYB  00000118263MO3  00106782MO48CS  00059632YY92CS  00000024578YY7  00000022987YY2  00000011106YY2  00000024544YY9  00000025721YY2  00000024529YY0  00000021111YY0  00000041701YY4  00000089565MO6  00059632YY25CS  00000147669MO6  00000024524YY1  00106782MO33CS  00059632YY67CS  00059632YY68CS  00000024619YY9  00000024617YY3  00000024618YY1  00000024616YY5  00000083259YY2  00000083258YY4  00000083253YY5  00000023043YY3  00000085055YY2  00000085063YY6  00000085087YY5  00000085129YY5  00000089512MO8  00059632YY63CS  00059632YY61CS  00059632YY69CS  00000021062YY5  00059632YY65CS  00000011098YY1  00000008113YY3  00000013783YY6  00000025737YY8  00000025771YY7  00000023002YY9  0106782MO105CS  00000025732YY9  00000059976YY1  00000089566MO4  00000089589MO6  00000106770MO1  00000155053MO2  00000140504MO2  00000010970YY2  00059632YY62CS  00000021117YY7  00000085033YY9  00000083279YY0  00000041774YY1  00000086012YY2  00000086007YY2  00000086014YY8  00000083283YY2  00000085035YY4  00000048255YY4  00000086015YY5  00000160482MO6  00000160479MO2  00000160475MO0  00000160483MO4  00000168891MO0  00000118290MO6",
      "center_classification_date": "20130410",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including the states of:  AR, CA, GA, IL, IN,IA,LA, MA, MN, MT, NY, OH, PA, TX, VA and the countries of BELGIUM, BURKINA FASO, BRAZIL, CHINA, CANADA, SWITZERLAND, CHILE, COLOMBIA, CYPRUS, CZECH REPUBLIC, GERMANY, DENMARK, DOMINICAN REPUBLIC, EGYPT, SPAIN, FRANCE, UNITED KINGDOM, CHANA, GUADELOUPE, GREECE, GUYANA, HONG KONG, CROATIA, HUNGARY, INDONESIA, ISRAEL, ITALY, KUWAIT, LEBANON LUXEMBURG, LIBYAN, MOROCCO, MARTINIQUE, MEXICO, NETHERLANDS, NORWAY, PANAMA, PERU, POLAND, PORTUGAL , REUNION, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SUDAN, SWEDEN, SINGAPORE, SLOVENIA, SLOVAKIA, SIERRA LEONE, THAILAND, TUNISIA, TURKEY, VENEZUELA, MACEDONIA, PALESTINE, SOUTH AFRICA.",
      "state": "WI",
      "product_description": "VMX, VMX Plus  Mobile Radiographic Unit with the following models:  45296139; 45296200; 45296350; 45296477; 45296964; 45297001; 45296966; 45297003; 2240169; 2240171; 2240172; 2240173.    Product Usage:  The VMX is a mobile radiography system incorporating stylish design for totally adapted use in intensive care, emergencies, pediatrics, neonatology and orthopedics.",
      "report_date": "20130417",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GE Healthcare, LLC",
      "recall_number": "Z-1083-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64759",
      "termination_date": "20140204",
      "more_code_info": "",
      "recall_initiation_date": "20120123",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}