  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "DuPont",
      "address_1": "2620 Williamson Place NW",
      "reason_for_recall": "Moisture level may cause the VisiPlugST (or CollaSyn Plugs) to break apart, or crumble upon contact.  The firm's data also suggests that a 4 year shelf  life may not be sustained.",
      "address_2": "Suite 113",
      "product_quantity": "781 units",
      "code_info": "LOT numbers:    072314-2328;   072314-2329;   072314-2330;   081314-2365;   081314-2366;   081314-2367;   091814-2345;  091814-2350;    100614-2344;     EXPIRATION DATES:    2018-05-15;  2018-07-23;  2018-08-13;  2018-08-14;  2018-09-14;  2018-09-14;  2018-09-25;  2018-10-06;",
      "center_classification_date": "20150210",
      "distribution_pattern": "Worldwide Distribution-US (nationwide) including the states of AR, CA, FL, ID, IL, KS, MA, MI, NC, NJ, NY, OR, PA, and TX and to the countries of Australia, Brazil, Czech Republic, France, Ireland, Mexico, Peru, Switzerland, and United Kingdom.",
      "state": "WA",
      "product_description": "VisiPlugST for the Lacrimal Efficiency Test.  It is also sold in the EU as CollaSyn Plugs.  Model number 1813, 1814, 1815, 1903, 1904, and 1905.    This device is cylindrical in shape, approximately 1.75-2.00mm in length, and available in three sizes: 0.3mm, 0.4mm and 0.5mm.    Plugs are place in the upper and/or lower horizontal canaliculus. Two plugs are placed into a foam carrier and nitrogen sealed inside a foil pouch. The foil pouch is sent to sterilizer for gamma sterilization. Two sterile pouches are packaged in a box with an IFU.",
      "report_date": "20150218",
      "classification": "Class III",
      "openfda": {},
      "recalling_firm": "Lacrimedics Inc",
      "recall_number": "Z-1081-2015",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "70036",
      "termination_date": "20150323",
      "recall_initiation_date": "20141126",
      "postal_code": "98327",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}