{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Waukesha",
      "address_1": "3000 N Grandview Blvd",
      "reason_for_recall": "GE Healthcare has recently become aware of a software issue associated with patient selection from the worklist on your WDR1(Precision 500D and Proteus XR/a equipped with Wireless Digital Radiography Option) that may impact patient safety.    Intermittently when a large number of patient exams are retrieved from the HIS/RIS system, the exams when displayed on the WDR1 are not properly displayed. I",
      "address_2": "",
      "product_quantity": "359 (333 US, 26 OUS).",
      "code_info": "Mfg Lot or Serial # System ID  00000007820HL1 757736WDR1  00000075975HL3 WDR1-RF1  00000075976HL1 530BGDR  00000075977HL9 210MTFP  00000075979HL5 337470WDR2  00000075981HL1 205408WDR1  00000075982HL9 WDR1-RF2  00000075983HL7 337470WDR1  00000075984HL5 LLPM500D  00000076011HL6 434654RM4  00000076012HL4 585596FPD1  00000076013HL2 520GV500D  00000076014HL0 781624HSSHFP  00000076015HL7 954JDCHWDR1  00000076016HL5 503391SKY  00000076018HL1 954JDCHWDR2  00000076019HL9 518926DRRM4  00000076021HL5 954265WDR1  00000076022HL3 207795CM3WDR  00000076023HL1 719285P500  00000076026HL4 540DALEVILLE  00000076027HL2 410392RAD  00000076028HL0 318495WDR1  00000076285HL6 0002952447  00000076286HL4 918579PROTEUS  00000076287HL2 518926FLRM3  00000076288HL0 ORMCP500DIG  00000076290HL6 713791P500FP4  00000076379HL7 518243WDR1  00000076380HL5 504873FP1  00000076381HL3 435722GDXR  00000076487HL8 989224FLSHPD1  00000076489HL4 229924P500  00000076490HL2 405752RF2  00000076643HL6 COR481360  00000076644HL4 713UTPROTEUS  00000076646HL9 518243WDR1  00000076647HL7 706754EF1  00000076648HL5 COR481360  00000076649HL3 518243WDR1  00000076650HL1 918273CNNP1  00000076651HL9 918434P500  00000076652HL7 714827P500  00000076986HL9 325574WDRRM2  00000076990HL1 878WDR  00000076991HL9 919RALCPFLASH2  00000076995HL0 912384FP5  00000076996HL8 540829RF1DR  00000076997HL6 305LEON27FLGD  00000076998HL4 205877BPRF1  00000077642HL7 989224FLSHPD1  00000077644HL3 240566EDWDR1  00000077647HL6 925370PAD  00000077648HL4 603524500D  00000077649HL2 618443SFP  00000077652HL6 540CULRF2  00000077653HL4 501202WDR14  00000078002HL3 760873FP2  00000078196HL3 318259RM2  00000078201HL1 757736WDR1  00000078669HL9 419538RAD  00000078670HL7 601928PXR  00000078671HL5 512LRMCRF  00000079045HL1 417347P500  00000079046HL9 608364PRFPCL  00000079047HL7 218828WDR1  00000079048HL5 269637FP  00000079049HL3 281274P5DGR  00000079457HL8 501202BNLER  00000079460HL2 507434DR1  00000079461HL0 570455GDXR  00000079662HL3 312CMP500D2  00000079663HL1 330841AND  00000080016HL9 724285WDR  00000080196HL9 325574WDRRM1  00000080556HL4 315425A320WDR1  00000080560HL6 301874UBWDR1  00000081032HL5 281477MWBP500D  00000081033HL3 412372DD1  00000081034HL1 608356P500  00000081035HL8 504896R2P500  00000081036HL6 252847EDDIG1  00000081037HL4 252847EDDIG2  00000081038HL2 504873FPR4  00000083548HL8 314919JDR1  00000083876HL3 732840WDROP  00000083877HL1 636239DR2  00000083878HL9 740383WDR1  00000083881HL8 636239MOBDR1  00000084098HL3 404778FPAD  00000084101HL5 0002996020  00000084206HL2 906293DIG2  00000084208HL8 304252FCPROT  00000084209HL6 518292RM4DR  00000084211HL2 781306FP  00000084313HL6 724983500D  00000084314HL4 904271FP1  00000084315HL1 206288WDR1  00000084551HL1 505609FLASH1  00000084552HL9 240566DF1  00000084553HL7 SLWPRO  00000084554HL5 330841PXD  00000084556HL0 330841RFD  00000084989HL3 409212GEDR1  00000085226HL9 828456FLSHPD1  00000085228HL5 732840WDRER1  00000085230HL1 641228WDR1  00000085231HL9 770400WDR1  00000085385HL3 412367WDR  00000085386HL1 815395OSFER1D  00000085387HL9 SLWP500  00000085389HL5 732321WDR3  00000085494HL3 330543WDR1  00000085496HL8 404265FPAD1  00000085604HL7 713SCFP  00000085605HL4 847872WDR1  00000085606HL2 970945ORTHO  00000085607HL0 925370PRO1  00000085609HL6 MH693WDR1  00000085927HL2 713SCFP2  00000086214HL4 731425TR3D  00000086216HL9 731425TR2D  00000086217HL7 636866DR1  00000086219HL3 908522RM8WDR1  00000086463HL7 508FHCMRAD  00000086464HL5 404501P500  00000086465HL2 507537WDR1  00000086466HL0 937OVFLWD  00000086629HL3 636239DR1  00000086631HL9 770968PRO  00000086632HL7 610861WDR1  00000086633HL5 318769ERM1  00000086635HL0 817347FP1  00000086969HL3 817347PRODIG  00000086970HL1 641446WDR1  00000086972HL7 708799PROT  00000087232HL5 404265FPAD2  00000087233HL3 COR481341  00000087379HL4 309344MRF1B  00000087380GK2 410362ER  00000087381HL0 718226PRM10WDR  00000087542HL7 908277WDR1  00000087543HL5 210703FP  00000087545HL0 240566RHRF1  00000087546HL8 706596SFER  00000087657HL3 303269FP1  00000087659HL9 856355P500  00000087660HL7 COR481342  00000087661HL5 651789WDR1  00000087698HL7 602406WD138  00000087701HL9 COMFORTFL3  00000087703HL5 COMFORTFL2  00000088105HL2 718226PRM1WDR  00000088107HL8 0002991687A  00000088108HL6 812934PFP1  00000088110HL2 412647WDR1  00000088396HL7 507434DR3  00000088398HL3 660885WDR1  00000088399HL1 936348MVPRO  00000088400HL7 941355P500D  00000088401HL5 606833P500D  00000089175HL4 760876FP  00000089180HL4 0004034682  00000089181HL2 508996HRAD  00000090002HL7 479587P59  00000090374HL0 507434DR2  00000090375HL7 0004014002A  00000090376HL5 760379FP  00000090626HL3 441236P500  00000090628HL9 865316WDR1  00000090629HL7 205333FL2  00000091206HL3 208422P500D  00000091657HL7 805418XR1  00000091659HL3 479587P52  00000091663HL5 440525PROTEUS  00000091664HL3 360346WDR1  00000091666HL8 479684P500  00000091668HL4 516796DIGDET  00000091878HL9 937208DET  00000091879HL7 505983ORTHPROT  00000091880HL5 719638DP  00000091881HL3 702304RF1DET  00000091882HL1 281450OPT220  00000092402HL7 508833SHRAD  00000092403HL5 281540NECCRF  00000092404HL3 512FCXR200  00000092660HL0 601968WDR2  00000092661HL8 301891P500  00000092662HL6 863314PROT  00000092663HL4 718670CLINIC  00000093036HL2 UICGERF1  00000093038HL8 518292RAD1DR  00000093039HL6 601968ERWDR  00000093040HL4 972899FMPRO  00000093310HL1 318448RF1FP  00000093311HL9 734712AABJ2  00000093312HL7 906341FP1  00000093314HL3 419636RAD  00000093707HL8 918286WDR1  00000093708HL6 512LWMOBRAD  00000093709HL4 732294P5D  00000093710HL2 817865FPSXR220A  00000093711HL0 229312MFP2  00000093712HL8 405744OSUVET  00000093960HL3 760873FP1  00000093961HL1 573686WDR  00000094228HL4 830703VVFP  00000094229HL2 937641RMF  00000094230HL0 818360XR1  00000094374HL6 904215RFWDR  00000094375HL3 330332R4WDR1  00000094747HL3 863314P500  00000094749HL9 207795P500WDR1  00000095000HL6 563589WDR1  00000095001HL4 906341DIG  00000095118HL6 229276FP2  00000095119HL4 719557WDRF1  00000095122HL8 812265WDR1  00000095380HL2 636789PDR  00000095381HL0 740633P500  00000095382HL8 608356P5002WDR1  00000095383HL6 718226VZRM3  00000095387HL7 718670CLINIC  00000095389HL3 573686MOB  00000095390HL1 229276FP1  00000095391HL9 978741SRF3W  00000096330HL6 732530URGW  00000096331HL4 570621WDR1  00000096332HL2 251435I7  00000096333HL0 218741WDR  00000096334HL8 806663RAD  00000096335HL5 601663P500  00000096511HL1 305MCHDORAL  00000096512HL9 314251MR2DR  00000096513HL7 505WHRM1  00000096518HL6 606433PROT  00000096635HL8 304234560  00000096636HL6 502484WDR1FLSH  00000096638HL2 513MWDIG1  00000096993HL1 973726WDR  00000096995HL6 718226VZRM2  00000096996HL4 719557WDRAD  00000098046HL6 717761GDXR  00000098047HL4 229353FP3  00000098051HL6 618664FCRAD  00000098052HL4 229276FP3  00000098053HL2 307672WDR1  00000098056HL5 941629WDR1  00000098057HL3 662751RF2FP  00000098058HL1 808YMP500D  00000098882HL4 513MWDIG2  00000098883HL2 801545FP  00000098884HL0 314251MD1DR  00000098887HL3 DCHRMADIG  00000098888HL1 904202ERDR  00000099152HL1 281420SJM500D1  00000099153HL9 636695MWR1  00000099391HL5 937734PRE  00000099401HL2 309672WDR1B  00000099402HL0 732840WDOP2  00000099403HL8 WDR1ED  00000099404HL6 309672P500D2  00000099405HL3 757889RM4  00000099949HL0 360805PROTEUS  00000099950HL8 815756WDR1  00000099952HL4 518489RF1WDR1  00000099953HL2 570621SEPREC  00000099954HL0 718630FPRM8  00000099955HL7 541276P500DR  000000Y0S11014 WDR1-RF2  000000Y0S11053 405752FP2  000000Y0S11056 660223WDR2  000000Y0S11062 918434P500  000000YOS11041 404686MOTDR1  00000100202HL1 478757DR  00000100203HL9 702255XR2  00000100206HL2 303399P5002  00000100207HL0 330725PROTEUS  00000100209HL6 229391WCFP  00000100211HL2 972566PEDIRF1  00000100385HL4 717761WDR2  00000100387HL0 308865KRDP  00000100388HL8 972841XR220A  00000100389HL6 954355UPGKIT  00000100390HL4 904202ER1DR  00000100714HL5 0004138298  00000100718HL6 281FCDPDM  00000100719HL4 513MWPRE1  00000100720HL2 717733R1B  00000100722HL8 FCLEWPROT  00000101087HL5 757261RM10  00000101089HL1 301877RAD3  00000101137HL8 MH693WDR2  00000101138HL6 310374PRO  00000101661HL7 0004114071  00000101663HL3 940384PRO  00000101669HL0 405372WDR1  00000101670HL8 405372WDR1  00000101671HL6 405949ER1  00000102207HL8 OVERPRO  00000102209HL4 218935WDR1  00000102210HL2 830393CFP  00000102211HL0 863680UCFP  00000102213HL6 415VASFP500  00000102214HL4 419294DR  00000102511HL3 601200FP1  00000102514HL7 304636WDR1  00000102515HL4 513MWPRE2  00000102516HL2 907276P500  00000102518HL8 818563DR  00000102891HL9 412647P500D2  00000102893HL5 412647F3WDR  00000102895HL0 210877PRO  00000102896HL8 281338P500  00000102897HL6 FCMESQXR220  00000102898HL4 0004074932B  00000103168HL1 503399WDR  00000103169HL9 208402RAD1  00000103170HL7 0002922151A  00000103171HL5 703697PROT1DR  00000103174HL9 724588PROTWDR1  00000103175HL6 0004074923B  00000103760HL5 518254WDR3  00000103761HL3 909CH500DFP  00000103763HL9 904321WDR1  00000103767HL0 570523WDRSIC  00000103975HL9 225658RF1  00000103976HL7 518254WDR2  00000103977HL5 0004076901  00000103979HL1 0004076901  00000104296HL9 417347FEP500  00000988882HL4 513MWDIG2  00000Y0S11007A WDR1-RF1  00000090630HL5 XR89403HL0  00000076989HL3 0910072218  00000083546HL2 0910072221  00000093959HL5 0910074035  00000095384HL4 83089910031  00000096515HL2 83098669808  00000096516HL0 83090416600  00000101667HL4 83093535827  00000080557HL2 613969RF1  00000094745HL7 416323RAD2  00000087234HL1 M1797703  00000101659HL1 IE1006WD01  00000085227HL7 A5805631  00000096517HL8 JO1034RX08  00000079664HL9 RO1096XR01  00000081031HL7 RO1096XR01  00000091661HL9 RU1279RX03  00000080195HL1 AE1021RX04  00000077641HL9 00141RAD05  00000080193HL6 10670WDR01  00000084555HL2 00313WDR01  00000084988HL5 00100WDR02  00000089178HL8 00618RAD01  00000094998HL2 00254RAD01  00000095388HL5 05528WDR01  00000101665HL8 IE1564RX03",
      "center_classification_date": "20140224",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including the states of  AL, AK, AZ, AR, CA, CO, DC, FL. GA, HI. IA, ID, IL, IN, LA, KY, ME, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, TN ,KY, TX, UT, VA, WA, WV, WI and WY., and the countries of   ARGENTINA, AUSTRALIA, CANADA, FRANCE, IRELAND, ITALY, JORDAN, ROMANIA, RUSSIA, UNITED ARAB EMIRATES and UNITED KINGDOM.",
      "state": "WI",
      "product_description": "GE Healthcare Precision 500D and Proteus XR/a equipped with Wireless Digital Radiography Option (WDR1).    The Wireless DR Imaging Option ( WDR1), when used with a radiographic imaging system, is indicted for use in general radiographic images of human anatomy. It is intended for use in all routine radiography exams, and specialized areas including pediatric work, intensive care and trauma, wherever conventional screen-film systems or CR systems may be used. The device is not intended for mammography or fluoroscopy applications.",
      "report_date": "20140305",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GE Healthcare, LLC",
      "recall_number": "Z-1078-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67401",
      "termination_date": "20141103",
      "more_code_info": "",
      "recall_initiation_date": "20140120",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}