  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Waukesha",
      "address_1": "3000 N Grandview Blvd",
      "reason_for_recall": "Accelerated fatigue of the lateral axis motion subsystem of the Infinia Computed Emission Tomography System could result in mechanical failure causing the detector to slide; may impact patient and operators.",
      "address_2": "",
      "product_quantity": "2180",
      "code_info": "System ID (As required)  787764INFINIA-1  SA2146NM01  TN1366NM01  SA1001MN01  PK1031NM01  PK1032NM01  PK1180NM08  PK1184NM01  PK1307NM01  PK1336NM01  MA1439NM01  LB1122NM02  KW1013NM01  KW1013NM02  KW1018MN06  KW1018MN08  EG1124MN02  323268INF1  630312INF  630545INF  630856INF  630275INF  330344INFINIA1  518262INF1  518262INF2  518262INF3  724226INF1  ARHINFI  610770INFINIA  502893INF  443481BINF  443481INF  443481INF2  661949NUC1  804594INF1  480323INFRM1  607737INF  404265INF  330363INFINIA  920451INF  920907INF  305682INF1  602588INF  904821INF3  904821NM1  904821NM2  615284INF1  904627INF2  904321INF1  904202INF1  501202INF  662244INFINIA  225381INF  972394INF  727863INF1  713359INF1  419483NUC2  406731INF1  228436INF  865977INF  804281INF1  804225INF1  941VRMCINF  248471INF  203384INF  269657SHINF  718240INF2  718240INF1  BCHINF1  513454INF1  843402INF2  843606INF1  609730INF1  GILESINF  217554CINF  843674IN1  843674IN2  843674IN3  704225INF1  704355INF1  704355INF2  704863INF1  770382INM  973877INFINIA  859313INF  423625INFINIA  859498INF1  937440INF2  606878INF1  606878INF2  512528NM1  305325INF  615342INF1  816380CASSINF  816392RINF  816836CIIN  SLGHINF  207795INFINIA  518562INF  757312INF1  757312INF2  410337INF3  973831INFINIA  708684INF1  302733INF  318681INFINIA2  281599INF1  805489INF1  352344INF  256327NUC  704487INFINIA1  718948INF  843782INF  931403NM1  256429INF1  CFRHINF  727521INF  904276INF1  702255INF  210617NM1  847990INF1  814533INF2  814534INF  607756INF  607756INF2  740623INFINIA  319272INF  CRAVENINF2  CRAVENINF3  CRAVENINF1  618544INF  402449INF  337436INF  318798INFINIA  203797INF  304637INF  205750INF  205759INF  256341NRM1  760778INF1  702369NM  760241INF1  850883ENM  904542NM  502624INF  727581INF  630786INF  786308INF1  786308INF2  214324INF  919684INF3  318330INF1  843412INF  504454INF  303837INF  330721INFINIA1  330721INFINIA2  502350INF  904819INF1  928214INF  FHEINF  954735INF2  813788INF1  225647INF  318327INF1  318388INF1  601288INF1  865373INF  865429INFINIA  240566INF  256494NM1  316858GI  662256INFINIA  561650INF1  843692IN1  319768INF  319768INF2  770229INF  979532INF1  706278INF  828586INFINIA  262670INF2  978521INF  517543INF  727942INF  256386HKNM2  906293INF  618942INF  305835INF  847480INF  843689INF  605353NM  435251INF1  801357INF1  701364INF  703858INF  815432INFINIA  814643INF  773947INF  570398INF  502955INF  859268INF1  502259INF  732321INFINIA  954265INF1  561744INF  305227INF1  305227INF2  262656INF2  513867INF  441236INFINIA  704739INFI  281348INF  307332INFINIA  225658INF  772LMCNM2  901287INF1  LGMCINF  304831LNM1  951672INF1  805737INF1  805370INF1  304623INF1  304623INF2  609835INFINIA  707822INFINIA  409938INF  828696INFINIA1  252436INF1  415925INF  925779IN  740383INF1  843431IN  402461INF  931381NM2  843777IN1  843777IN2  814333INFINIA1  814333INFINIA2  478633INFINIA  732840INFINIA  580931INF1  727843INF1  815756INF  609267INF  713242INF2  713242INF3  432685DNUC  812522INF  MVINFNM2  913345INF1  209385IN  419455INF1  701774NM  281274INF  402354INF  832522INF  832522INF2  614478INF1  615349NM1  615349NM2  860344INF  662772INFRM1  662772INFRM2  662772INFRM4  405610MINF1  617313INF1  603924INF  973971INFINIA  212241INF2  801465INF1  302629INF  NASHNM2  601443INF1  518773INF  239NCHINF1  281897INF1  212608INF1  COR373273  COR373282  617243INFINIA  405307INF1  352333INF  225NOINF  985230INF  770219INF  770535INF1  210757NM1  215537INFINIA  985646INF  479751INF  954978INF1  203852INF  704316INFI  516IPKINF1  352596INF1  708660INF  734467INF  734671INF  228818INF  885INFI1  885INFI2  478454INF  801479INF  801479INF2  ORMCINFINIA2  OBNM1  913541INF1  908522INFINIA2  908522INFINIA3  417256INF  661PMCNUC1  845987INFINIA  435658INF  423493PINF  803329INFIN1  PITTINF5  571231INF1  571231INF2  610327INFINIA  303839INF  815432INF  248849INF1  248849INF2  903785INF  505863INF  816348INF1  804254INF  618395INF  601936INF1  732530INFINIA  256543INF1  325235NUC  843402INF  843724INF2  773995INF  828286INFINIA  920849INF  918494INF2  918494INF3  918502INF1  501776INF  518271INF  310832INF1  502868INF1  480882INF2  435893INF  951696INF1  818907INF1  805543INF1  479524INF  605333INF2  985649INF1  985649INF2  856363INFINIA  404466INF  229259INF  229333ICINF1  229333ICINF2  912691SCNC  918294INF1  610869INF  615781NM1  301877INF  740356INF2  931967NM  804957INF1  804957INF2  616267INFINIA  540498INF1  828766INFINIA  504537INF  815395INF  309344SMNUC  217347INF2  217347INFINIA  985785INF  412343INF1  618234INF1  918307INF1  901765INFRM8  864675INFI  516562INFINIA2  309655INF1  708747INFACQ  225INFINIA  313343INF2  313343INF3  313343INF4  217544NUC1  773665INF1  773665INF2  765456INF  813870INF  410337INF2  602406INF  602406OPINF  281251SLVINF  713785SLINF2  610954INFINIA  281637INF  479967INF  580249IN1  409985NMI1  812485INF  214879INF  814452INF1  205408INF2  317415INF  912486EGNC  405742INF1  615328NM1  504380INF  504873INF2  404686INF  573686INF2  662423INFINIA  828883INFINIA  734475INF  903731INFINIA  918307INF2  570836INFINIA  310319INF1  806354NWINF  330343INFINIA  704225INF2  724450INFINIA  607762INF  681342INF1  681342INF2  681342INF3  681342INF4  434924INF1  434924INF2  434924INF3  913588INF1  774442INF  618256INF1  817472INF  208422INF  254743NM2  713794INF3  541440INF  269223VABCINF  906774INF  618993INF  757722INF1  509473INFINIA  702388VNM  910323INF1  970945INF  706733INF  478272INF  419238INF  210MTINFINIA  316268VCFI2  814723INF  479463INF  479463INF2  828262INFINIA  856247INF  623882INF  620225WPI  262687INF2  248898INF  248898INF2  757259INF1  814467INF  713722INF  817346INF  512324NM  210705NM1  765552INF  740264INF4  740264INF3  630275INF2  818884INF1  UK5002NM01  0853260023  0853260066  0853260074  0853260046  0853260056  0853260039  0853260087  0853260024  0853260045  0853260069  0853260042  0853260072  0853260062  0853260020  44014NUC01  06109NUC01  34318NUC01  01012NUC01  17005NUC01  21011NUC01  35003NUC04  23003NUC01  34284NUC04  45011NUC01  34335NUC01  23002NUC01  06008NUC09  06051NUC01  41009NUC01  52006NUC01  34017NUC01  07015NUC01  34094NUC02  63009NUC01  38001NUC01  38001NUC02  00013NUC01  01042NUC01  34021NUC02  34340NUC01  34351NUC01  0856260029  0856260031  0856260026  0856260028  0847260025  0002765581  A004NU11  Y004NU11  PPR12213  PPR59030  PMR12506  PPS06907  PK1001NM01  0843260034  0843260037  0843260044  2565059INFINIA  NO1000MN06  HGV0130  GOU0137  GOU0138  HDB0130  MPP0131  MT1014NM01  MT1014NM02  MQ1000MN04  LT4000MN01  0850260077  0850260088  0850260089  0850260106  0850260107  0850260112  0850260079  0850260060  0850260061  0850260087  0850260076  0850260090  0850260105  0850260110  0850260068  0850260121  0850260065  0850260103  0850260081  0850260115  0850260123  0850260120  0850260093  0850260094  0850260063  0850260069  0850260070  0850260067  0850260073  0850260124  0850260091  0850260095  0850260104  0850260114  0850260083  0850260092  0850260113  0850260066  0850260080  0850260082  0850260099  0850260064  0850260100  YN1562  YN1566  YN1573  YN1581  YN1584  YN1571  YN1559  YN1579  YN1570  YN1551  YN1588  YN1575  YN1543  YN1586  YN4930  YN1567  YN1552  YN1561  YN8003  YN1560  YN1545  YN1554  YN4949  YN8001  YN1583  YN1556  YN1546  YN1558  YN1574  YN1544  YN1568  YN1569  YN1557  YN4922  YN1542  YN1563  YN4923  YN5000  YN1549  YN1564  YN1565  YN1547  YN4929  YN1541  YN1582  YN1540  YN1539  YN1550  YN4990  YN1577  YN1548  YN1587  YN1589  YN1578  YN1553  YN1196  YN1580  YN1555  YN1572  YN1585  YN4920  YN4918  YN4919  YN4924  YN4925  YN4927  YN4934  YN4921  YN4931  A5279302  A5181662  A5634003  B5142302  B5142304  A5315551  A5113627  A5178021  B5395703  C5120912  A5924402  A5109665  A5109666  A5216824  A5162248  A5352208  D5111607  A58080111  A58080134  A5111240  A5215935  A5129026  A5267107  A51043155  A51043187  A51152163  A51152164  B5365422  083026602101669  083026802101686  083026801003210  083026602101279  083026002002289  IL1069NM06  IL1068NM05  IL1018NM07  IE1006MN04  HU1099NM01  HR4103NM01  05452NUC03  00271NUC04  00624NUC01  00624NUC03  00116NUC01  00186NUC02  IE1564NM03  00289NUC02  00203NUC03  00231NUC09  00165NUC04  IE1510MN09  00669NUC01  00169NUC07  10497NUC01  00440NUC11  00440NUC07  70027NUC03  00171NUC04  77624NUC01  77624NUC03  V1471114  V2943415  V7406615  V7405209  M4159227  M4159228  M4018536  M4163763  M4018099  B5483443  A4169304  M8572211  M1400803  M4024095  A4194407  X105967813  M2269829  V2728105  M4167153  M218669001  M4211068  M5172527  X61014001  X100611101  M4023087  M9886832  M1279727  M4480232  M1804913  M4135141  B4182004  M4234232  M4234235  M4479619  M44827146  E4858901  M4144332  P4211482  R4211480  M9162716  X59007307  M7411106  X2034002  X151957601  S4160059  M4040544  M4215553  X4211026  FE443890  DE539254  UEW19003  UEW19005  MPX46902  DEN01721  DEN01727  UPW00402  GPQ19105  ME051533  UE720407  MPX82204  MPX47106  EG1826MN01  DK1023NM05  HC7475NU03  HC0842NU01  CHAM01NU01  UC7655NU03  HC0507NU03  HC1249NU01  HC1249NU02  493519NU01  PC1818NU03  PC8118NU01  HC1819NU01  CC3899NU01  HC3442NU01  HC0892NU04  HC4387NU02  2583615INF  2928217inf  262010NU01  475290NU01  529150NU04  905683IN1  613969INFINIA  819537INFINIA  819537INFINIA2  416928INFINIA  709737INFINIA  204837IN1  416813INFINIA  416444INFINIA1  905855INFINIA2  905855INFINIA3  905472INFINIA  905832INFINIA  613727INFIN  514934AINFINIA  416756INFINIA2  905773INFINIA1  905845INFINIA1  705759INFINIA2  403777INFINIA3  306655INFINIA2  306655INFINIA4  905895INFINIA1  905895INFINIA2  905522INFINIA  905522INFINIA2  705674INFINIA  519965INFINIA  905848INFINIA  250727INFINIA1  082426040034  082426040035  082426270006  082426190016  082426140020  082426140021  082426120006  082426120017  082426040037  082426110007  082426140017  082426150022  082426040033  082426070022  082426240003  2883051  NMINDTCRL0311  2827087  080004NU13  160150NU04  030018NU05  110093NU07  030010NU05  030037NU04  080053NU02  100029NU01  010120NU11  130083NU13  190073NU01  0910262234  0910262175  0910265053  0910265059  0910264085  0910264058  0910265055  0910264062  0910263190  0910264079  0910264129  0910262196  0910264152  AH5689NU01  AH3403NU01  ZA1820NM04  215952INF2  083026008153639  083026202100929  083026002000469  083026822101249  083026002100219  A5107766  DZ1338NM01  KW1005MN11  KW1018MN05  JO1034NM01  EG1100MN01  EG1538MN01  DZ1035NM03  DZ1035NM04  PK1180NM03  PK1335NM01  MA1272NM01  KW1005MN14  DZ1279NM01  412653INF1  510437IN  650725IN2  RU1152NM02  YN4252  YN4821  M256517203  876259NU01  CY1015NM01  849010NU01  0910262335  06109NUC02  06014NUC02  34329NUC01  35003NUC05  34017NUC02  RU1284NM01  RU2050NM01  RU2527NM03  RU3424NM01  PPR97306  LV4002NM02  A5197934  B5119121  F5119151  B5174320  IR1049NM01  00105NUC03  00393NUC06  00131NUC03  00133NUC21  10448NUC04  00373NUC01  V2147210  V7406617  M4159235  V6414104  B1650218  NPV15964  MPX52309  DE492A51  MEX68919  YE364417  DK1065NM05  HLUE01NU01  493519NU02  662608NU01  CS1021NM03  CS1318NM01  110064NU01  150017NU05  190171NU05  110009NU01  M23084315  ZA2261NM01  701234INF2  MPX99201  414384INF  A5719812  M4145143  M54606103  M40345193  0856260023  0834260003  26329NM1  2962725INFINIA  3882942310MNCTR  0834260008  2605823INF  0835260013  941235INF  763MRCYINFI  763UNTYINFI  920738INF  630801INF  732292INFINIA  623478NM2  214820INF1  972498INF  941766INF2  716366INF  973625INFINIA  865545INFINIA1  937278INF  601376INF1  941637INF  865947INFINIA1  614722INF1  302428INF  318681INFINIA  626359INF1  626359INF2  518828INF  972969INF  717627INFINIA  814533INF  812842INF1  906776INF  404712INF  781979INF1  615768SCNM1  731644INF  507345INF1  919934INF  847535INFACQ  863679INF1  863687INF1  863687INF2  717544INFINIA  801408INF  989723INF2  937328INF2  860350INIF  704783INFI1  704783INFI2  704783INFI3  937208INF2  908522INFINIA  717782INF  PITTINF4  704384INFI2  704384INFI5  704384HAWK1  262363INF  661633NM1  585922INF1  843724INF  715839INF  858554INF2  605333INFINIA  740356INF1  210921NM1  415222IN1  205408INF  203325INF  203325INF2  504873INF  573686INF1  907562NM1  540213INF1  601978INF  956350NM1  316962WI  01002NUC04  0856260025  O005NU14  Y004NU08  H002NU01  PPR08202  0843260029  2543334INFINIA  YN1200  A5153612  A5197927  C5120915  A5172907  A5118056  083026862101014  083026602101643  083026862100874  083026822100376  IE1100MN01  IE1029MN05  IE1053MN01  HU1039NM02  0828260003  0828260005  00109NUC01  00253NUC02  00216NUC11  00260NUC03  00348NUC02  IE1510MN12  00331NUC02  UE282633  YE364413  783348NU01  450654INF2  709651INFINIA  604520INFINIA1  604520INFINIA2  519749INFINIA  418275INFINIA  905731INFINIA  905855INFINIA  416633INFINIA  416756INFINIA  905372INFINIA  705759INFINIA  306766INF2  306766INF1  306655INFINIA  519631INFINIA  604588INFINIA1  604588INFINIA2  604278INFINIA2  604875INFINIA1  082426040030  082426210013  082426120011  082426200003  082426150012  082426130025  082426130020  082426160025  082426080010  082426100028  082426030020  082426120007  082426030021  082426130024  082426080012  082426130021  082426310015  082426160026  082426050004  082426030022  082426010014  082426140016  082426070018  082426110004  2812777  110034NU04  408259IN1  508828INF  815741INF1  815741INF2  970350IH  757398INF1  916537IN  415677INF  619482INFINIA  941629INF  530876IN  414805QUA  262574INFIN  408559IN1  608775INF1  608775INF2  262670INF  928855INF  920684INF  610437INFINIA  925INF  509946INF  310543INF1  602344NM1  603650INFINIA1  623848NM2  409727NM1  989723INF1  937328INF1  215952INF  516570INF  501620INF  831755IN1  928537INF  937208INF1  225765INF1  704384INFI4  405271IH1  801588INF  516622INFINIA  303320IH  509474INFINIA  626570INF1  858554EDCINF  858554INF1  757668INF1  650367IN1  480301NM1  734246INF  314525INF1  713785SLINF1  817882INF1  410955INF  205939INF  907562NM2  714456INFINIA  608263INF  608263INF2  650725IN  801584INF1  801584INF2  415750IN1  305575INF1  909825INFINIA  540536INF1  615322NM3  410605INF  206223INF  0856260024  0843260032  083026602100924  519752INFINIA  450654INF1  604520INFINIA3H  514842AINFINIA  604588INFINIA3H  604588INFINIA4H  604278INFINIA1  604875INFINIA2H  082426220007  082426200004  082426210015  082426270005  082426070019  082426190014  082426160027  082426180009  082426070020  082426120013  082426130026  082426180007  0910262170  0910266042  0910267010  MA1020NM01  KW1001MN23  PK1278NM01  IR1501NM01  UK1110NM01  RU2526NM01  RU2199NM01  RU2527NM02  NO1016MN16  NO1068MN07  DVT0138  LDD0431  LDD0433  HGL0233  HLV0334  A5367805  A5112411  A5120724  A5109126  A5198016  A5111127  A5115111  A5189714  IL1020NM25  IE1029MN04  GE1004NM01  10181NUC06  00109NUC04  00311NUC01  01367NUC01  00361NUC05  00139NUC05  00137NUC01  01243NUC02  00379NUC01  00318NUC03  70060NUC12  00187NUC01  00440NUC09  00100NUC02  77137NUC01  77139NUC05  M2846905  C4191749  FI1153MN01  FI1019MN01  FI1003MN06  FI1003MN05  DPN00002  NP739716  CE561221  FPG86405  HC0074NU01  PC5995NU01  485520NU01  CS1277NM02  900010NU06  110022NU03  300013NU01  030209NU02  AH5689NU04  ZA1807NM11  ZA1806NM12  ZA2008NM04  ZA1808NM13  ZA1835NM24  ZA2006NM04  505262INF  651241INF1  619229INF  781477INF  706651NUC  262948INF  507434INF1  256761INF  727893INF  302744INF1  757398INF2  518626INF  956687INF2  419227I1  419227I2  VASINF  804323INF1  775423INF1  361572NM1  281338INF1  915521HNM1  503435INF  845794INFINIA  6525011INF1  925370IN  361985SNM  406238INF  812426INF  757889INF1  719333IH  207255INFINIA  678474INF  703698INF1  626573INF1  818409INF1  806GRACENM  937374INF1  706494HNM  801501INF1  801314INF1  801IMCINF1  925785IN  410550INF  704660HAWK  262741INF  LGMCINF2  435750INF2  510400IN  713242INF1  713ACCINF1  228865HWK  262767INF  916983INFN1  530225IN  937390INF1  218333INF1  701234INF  626898INF1  727461VG  212241INF3  402559INF1  760873NUC  504838INFH  315592INF  704384INH1  915577NM1  706291HINM  956971NM1  802775INFINIA  503561INF  757388INF2  913676SMIM  508674INFINIA  608791INF  918494INF1  707963IN  410337INF  262334INF  901495INFHAWK  414489INF  801268INF1  209526IN  714953INFINIA  303399IH  254743NM  404321INF  304264INF  850494HAWK  928442INF  AC001NU14  600042NM05  NO1088MN09  ZWL0437  ALM0131  0850260074  0850260075  YN1601  YN1590  YN1597  YN1599  YN1592  YN1600  YN1596  YN1591  YN1595  YN1593  YN1598  YN1594  A5279301  A51149103  A5306502  D5111602  083026003811611  083026822100655  IL1069NM11  IL1331NM01  0835260010  0835260007  0828260006  00371NUC04  00130NUC03  00136NUC03  00234NUC05  00318NUC10  10497NUC03  M9164758  M4135145  B4155415  M4198189  M9319116  FI1000MN04  FI1064MN01  EEF51540  DK1026NM08  HC0863NU01  403382INFINIA2  613969HAWKEYE  418589INFINIA  819595INFINIA  905883HAWKEYE  418649NUC1  905813INFINIA  403289NUC1  604434INFINIA  250785INF1  514495INFINIA  819357BINFINIA  519685INFINIA  604531INFINIAH  506739INFINIA  306655INFINIA5  705674HAWKEYE  709637INFINIA1  082426290005  2638808  190057NU05  190187NU01  080026NU08  210036NU06  190205NU06  0910263143  0910263161  0910263166  0910263157  0910265050  0910266053  0910263159  AH5293NM04  AH5249NM04  PK1355NM01  KW1001MN25  KW1006NM01  45001NUC01  615896NM1  507402INF1  920456INF  814889INFINIA  480323INFRM3  435781INF1  540332INF2  540332INF1  262329INF  262434INF1  262434INF2  918376INF1  205783INF1  920288INF  217223INF1  217223INF2  814362INFINIA  802257INF  718250INF1  617296INF  937440INF1  916537IN2  607274INF  607274INF2  702360CHNM  423553INF  212989INF1  559225IN  315425INF  412692INF1  412692INF2  513585INF  361881SSINF1  785462CI  423472INF  972540INF  512397INF  507457INF1  414649INF  212686INF1  804675INF1  480512INF1  901523INF1  901523INF2  571231HAWK1  423778EDINF  423778MMINF  954735INF1  615327VANM1  615893NM1  920568INF  828315HWK4  865271NM1  518926INF  937832INF1  508427INF2  602239INF1  717633INF1  717633INF2  570501INF  781340INF  781340INF2  781340INF3  816271SJINF  801ALTAINF2  801IMCCARD  801IMCINF  801SLCINF  801357INF2  701857NM  724357INF1  703858HAWK  626915INF1  731425INFH1  731425NMIC1  731425NMRM  701952NM  215728INFINIAJ  732776INFH4  732321IH4  JOHNINF  408851IN  415838IN  4082IN  262656INF1  262741INF2  801299INF1  801408INF2  610402INFINIA2  718150INF1  6502IN1  540741INF1  NM156449  801387INF1  478633INFINIA2  972596INF  337981INF  713456INF  410822HK1  610278INFINIA  916453IN  608756INF  901516NINFHK  254519NMINF  541296HK4  718405INF1  231935INF  615769NM1  COR359309  COR369576  770751INF  817255INF  662377INFINIA  937438INF  901516UINFHK  718334INF1  718334INF2  718963INF1WH  718616INF  507OLMINF1  ORMCINFINIA1  337289INF  270688INF  404605INF  803329HAWK  802388INF  570621INF  425261INF  541387INF  248465INF  360PSPHINF1  703689INF4  919784INF1  401444INF  801RHINF  732937INFINIA  540689INF1  540689INF2  503561HK4  281420INF  505609INFINIA  408315IN  480882INF1  479441INF1  864SPAHK1  864SPAHK2  864560HAWK1  217528NUC  314525INF2  215427INFINIA  920738INF1  920738INF2  864675HAWK  516562INFINIA1  757886INF1  316268VCFI  313343INF1  484503INFINIA1  215710INFINIA  317338HAWK  540741HAWK1  228523HWK  916782IN1  916782IN2  215707INFINIA  832824INF  816404INF1  920729INF  607763INF  202715INF4  210358IHE  916703IN2  312996INF  913588INF  913588INF2  903UTINF  419383NM1  832505NM1  214370INF  813USFINF1  828298HAWK1  828298HAWK2  650858IN1  650858IN3  801584INF3  937268INF1  520VAINF1  713794INF  713794INF2  562826INFINIA  718630INF1  305575INF2  559353IN  540536OCHAWK  VVMCINF1  973676INFINIA  502287INF  VCWESTINF  757395INF1  206223INF2  301714INF  863680INF1  513298INF1  513298INF2  623207INF  262687INF  414325INF  570552INF  586416INF  203688INF  0853260040  0853260025  0853260036  0853260032  0853260030  0853260035  0853260029  06051NUC04  0856260030  0856260032  C001NU06  C001NU10  N001NU02  N001NU06  AC001NU18  RU3314NM01  RU1018NM01  RU3508NM01  PER03520  PER08009  787653INFINIA2  PL1143NM06  PL1692NM01  600042NM06  0920269087  0920269092  0920269093  NO1068MN09  APL0232  ZTP0133  EMM0131  0834260012  2750751INF  NMCR-17849  2766055INFHE4  36INF2753220  0850260108  0850260111  0850260096  0850260119  0850260117  0850260098  YN4237  YN4225  YN4248  YN4245  YN4243  YN4288  YN4223  YN4224  YN4239  YN4241  YN4242  YN4255  YN4236  YN4229  YN4300  YN4221  YN4250  YN4238  YN4228  YN4231  YN4226  YN4240  YN4234  YN4235  YN4222  YN4230  YN4244  YN4220  YN4219  YN4218  YN4227  YN4232  YN4233  YN4260  YN4822  A5127292  B5618331  A99263311  A186363001  A5123523  A5128755  A5129959  B5315345  A5118088  083026202101361  083026022002299  083026062102318  083026842101377  083026202100886  083026802100537  083026867892981  083026600020711  083026002101088  083026022002291  IL1043NM20  IL1062NM26  IL1059NM07  IL1020NM31  IL1018NM08  IL1041NM18  IL1041NM22  IE1110MN01  IE1085MN05  0828260015  0828260011  0828260009  0828260019  00002NUC02  00141NUC06  00065NUC05  00197NUC11  00433NUC09  00487NUC04  00374NUC05  00257NUC06  00257NUC15  00018NUC04  IE1510MN16  00240NUC05  00053NUC09  00383NUC03  V2943412  M4159460  M4160449  M4163034  M4111216  X105967809  V2728109  H4858974  M4194031  M4480229  B1768001  C4191765  M4184856  M2645705  M204493801  M4191430  M4144340  M1106608  M4143953  M7411101  X119197203  V8572204  S4160079  M4496943  FI1010NM01  FI1029MN01  EE1019NM02  ME208350  UE282648  ME244391  ME244393  YE626937  FEG46134  DE6588112  EE238526  UP285A41  DEN01723  DEN01725  DEH13556  DE528467  FEE12426  NE209938  FEB33835  18076INFINIA  DK1065NM04  DK1023NM03  HC2720NU02  PC3970NU02  485520NU07  PC1268NU03  PC2261NU02  HC4480NU01  PC3964NU02  600044NM07  CS1052NM08  600131NM05  600055NM04  CS1008NM03  260010NU05  529150NU03  722050NU01  403251INFINIA1  403382INFINIA1  403670NUC8  519752INFINIAH4  450654INF3  450468INFINIA  514371INFINIA  819378INFINIA  902567NUCINH4  709651INHE4A  403541INFINIA  905813INFINIA2  403289NUC2  709737INFINIA2  709737INFINIA3  604557INFINIAH1  604557INFINIAH2  604557INFINIAH3  604520INFINIA4H  905521INFHE4  905577INFHE4A  905577INFHE4B  204787INFINIA2  514252AINFINIA1  514252AINFINIA2  514252AINFINIA4  514252AINFINIA6  514842AINFINIA2  450746INFINIA  780489INFINIAH4  780489INHE4B  780489INHE4C  519685INFH4  705759HAWKEYE  403777INFINIA  403777INFINIA2  306655INFINIA3  514340INFINIA  250339INFINIA4H  416864INFINIA1  416864INFINIA2  416469INHE4A  416469INHE4B  250565NUC4  519663INFINIA2  604228INFINIA1  250370INFINIAH4  082426040039  082426210016  082426160034  082426200007  082426100033  082426100037  082426190015  082426150024  082426090018  082426110009  2768147  2694593  NMINDTCRL0227  2621801  1153389  030240NU07  030078NU13  190057NU10  030419NU09  190282NU09  030232NU10  030151NU12  190052NU06  190194NU01  030388NU04  0910262236  0910262288  0910262291  0910262210  0910262269  0910262307  0910262217  0910262276  0910263198  0910265066  0910264081  0910264089  0910264102  0910264112  0910264123  0910266050  0910266067  0910266063  0910267015  0910267020  0910262263  0910263210  0910263135  0910262255  0910265068  0910264072  0910264095  0910264121  0910264077  0910264083  0910264105  0910264107  0910264126  0910262312  0910263204  0910262205  0910262214  0910262220  0910262295  0910264075  0910264091  0910262239  0910262229  0910262241  0910262225  0910262227  0910262249  0910264100  0910263207  0910262244  0910262280  0910262301  0910262257  0910262285  0910265063  0910262293  0910264115  0910262266  0910262279  0910262332  0910264137  0910263220  0910265072  AH5689NU03  AH5117NU02  AH1908NU01  AH4660NM06  AH4307NM05  2954INFINIA  ZA2246NM01  ZA2212NM02  2723INFINIA  818409INF2  IL1041NM28  IL1041NM29  IL1042NM09  1577INFINIA  614722INF2  319356INF  809758INFINIA  PL1143NM03  00053NUC06  FI1021MN01  FI1023MN02  DK1029NM05  DK1022NM05  CS1277NM01  259170NU02  082426040041  082426090016  082426250003  082426120015  030232NU13  919684INF1  082426040036  082426100036  082426150013  082426150015  082426130023  082426160030  082426040038  082426150019  082426080014  082426150016  082426150018  082426120016  082426150014  415300IN  513636INF  402559INF2  916781IN  415353IN  916703IN1  787758INFINIA2  78NMIH2564713  0920269081  0920269079  NO1000MN03  NO1070MN07  AMS0338  INF2649553  2660585INF  2633201INF  YN1744  YN1743  YN1745  YN1742  A5547414  A5152964  A5114292  A5107756  B5357630  083026202100317  083026842101817  IL1028NM26  0828260007  IE1564NM01  2612588INF2H  DK1026NM07  427459NU02  2585033INF  519685INFINIA2  506544NUCGE  2775945  2555007  AH3701NM05  A5172908  209831IN  671USNHINF  903416INF  YN1650  YN1641  A5649008  A5216823  A5128757  A5124139  H2100NUC02  080004NU11  A5224207  919684INF2  650858IN2  600092NM10  PL1003NM03  4716INFINIAHE4  YN4830  YN4900  YN4820  YN4249  YN4819  YN4818  YN4825  083026006866397  083026002000938  2721482INF  082426100035  082426270007  082426050005  082426090022  082426190017  082426120018  082426230016  082426160035  082426100032  082426040042  082426170003  082426140022  082426070025  082426210017  082426120019  082426110008  082426290006  082426140019  082426230015  082426070023  082426100038  082426090024  082426140018  082426100041  082426090017  082426160033  082426090020  082426160031  082426310022  082426100043  082426130027  082426090021  082426080019  082426110006  082426070024  082426080018  082426310021  1299296  2769200  AH5844NM05  519749INFINIA2  082426310016  082426120021  082426130038  H010NU04  E001NU16  NO1054MN01  NO1116MN09  DHG0539  A5181660  00439NUC08  220826NU01  ZA1835NM25  703558INF  214820INF2  718584INFINIA  610402INFINIA  402354HWK  920272INF2  713785SLINF3  0920269082  2864298INF  0850260109  0850260084  0828260017  0828260018  0828260013  2694427INF  2646005INF  819563CINFINIA  403229INFINIA  905358INFINIA  204787INFINIA  250754INFINIA  250754INFINIA2  613761INFINIA  082426150021  082426160032  2760723  0910264067  0910264069  0910264098  0910266045  0910264060  787848INFINIA  773884INF  210575INF1",
      "center_classification_date": "20130410",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including DC and  PR  and the countries of UKRAINE, TAIWAN, TURKEY, TUNISIA, THAILAND, SINGAPORE, SWEDEN, SAUDI ARABIA, RUSSIAN FEDERATION, PORTUGAL, POLAND, PAKISTAN, PHILIPPINES, PANAMA, NEW ZEALAND, NORWAY NETHERLANDS, MALAYSIA,  MEXICO, MALTA, MARTINIQUE, MOROCCO, LATVIA, LITHUANIA, LEBANON, KUWAIT, KOREA, JAPAN, JORDAN, ITALY, IRAN, INDIA, ISRAEL, IRELAND, INDONESIA, HUNGARY, CORATIA, HONG KONG, GREECE, GEORGIA, UNITED KINGDOM, FRANCE, FINLAND, SPAIN, EGYPT, ECUADOR, ALGERIA, DOMINICAN REPUBLIC, DENMARK, GERMANY, CZECH REPUBLIC, CYPRUS, COLUMBIA, CHILE, SWITZERLAND, CANADA, CHINA, BRAZIL, BELGIUM, AUSTRALIA, AUSTRIA, ARGENTINA, SOUTH AFRICA, VENEZUELA, URUGUAY, ALGERIA.",
      "state": "WI",
      "product_description": "GE Healthcare Quasar Nuclear Medicine System, Hawkeye Option and Hawkeye 4 Option for Dual-Head Variable Angle Gamma Camera.  Nuclear Medicine imaging systems.  Recall includes all Infinia systems, all configurations.",
      "report_date": "20130417",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GE Healthcare, LLC",
      "recall_number": "Z-1078-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64398",
      "termination_date": "20140127",
      "more_code_info": "",
      "recall_initiation_date": "20120306",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}