  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "700 Orthopaedic Dr",
      "reason_for_recall": "Knee implant femoral devices were found to have anomalous microstructure. Porosity and the presence of blocky carbides, within the cobalt chrome part, can decrease the overall fatigue strength. This decrease in strength could cause implants to fracture under what would normally be acceptable loads.",
      "address_2": "",
      "product_quantity": "2",
      "code_info": "Catalog Number 960015  Lot Numbers 7810268 and 7806929 (US)  Lot Number 7806934 (OUS)",
      "center_classification_date": "20140224",
      "distribution_pattern": "Worldwide Distribution-USA including the states of IA, CA, KY, and FL, and the countries of Finland, Germany, Sweden, and Czech Republic.",
      "state": "IN",
      "product_description": "PFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5/RT)    Used during primary total knee arthroplasty to improve patient mobility.",
      "report_date": "20140305",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "DePuy Orthopaedics, Inc.",
      "recall_number": "Z-1077-2014",
      "initial_firm_notification": "Telephone",
      "product_type": "Devices",
      "event_id": "67159",
      "termination_date": "20140610",
      "more_code_info": "",
      "recall_initiation_date": "20131209",
      "postal_code": "46582-3994",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}