  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 Gbc Dr Ms 514",
      "reason_for_recall": "Complaints were received regarding Quality Control (QC) shifts with Pseudocholinesterase (PCHE) and imprecision during PCHE calibration with the Dimension Vista Enzyme 1 Calibrator (ENZ 1 CAL).",
      "address_2": "PO BOX 6101",
      "product_quantity": "3496",
      "code_info": "lot 3FD034, exp. 6/17/2014, and lot 3HD012, exp. 9/1/2014.",
      "center_classification_date": "20140224",
      "distribution_pattern": "Nationwide Distribution.",
      "state": "DE",
      "product_description": "Dimension Vista Enzyme 1 Calibrator (ENZ 1 CAL) lots 3FD034 and 3HD012.    An in vitro diagnostic multi-analyte calibrator for the calibration of Amylase (AMY), Gamma-Glutamyl Transferase (GGT), Lactate Dehydrogenase (LDH), Lipase (LIPL), and Pseudocholinesterase (PCHE) methods on the Dimension Vista System.",
      "report_date": "20140305",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-1076-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67235",
      "termination_date": "20150506",
      "more_code_info": "",
      "recall_initiation_date": "20140106",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}