{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72979",
      "recalling_firm": "Ortho-Clinical Diagnostics",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "N/A",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide including Puerto Rico, and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany,  India, Italy, Japan, Mexico, Panama, Singapore, Spain, and  Venezuela",
      "recall_number": "Z-1071-2016",
      "product_description": "VITROS 250 Chemistry Systems, Catalog 8132086, Unique Device Identifier No. 10758750004409, and VITROS 250 Refurbished, Catalog 6801759, Unique Device Identifier No. 10758750001330; IVD.    Product Usage:  For in vitro diagnostic use. The VITROS 250 Chemistry System performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides.",
      "product_quantity": "US: 802; Foreign: 2472",
      "reason_for_recall": "Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chemistry Systems when using Calibrator Kit 9, Lot 954.  A trend of complaints regarding customers actions following U90-382 or 6LU condition codes was noted (See RES 72289 - VITROS Calibrator Kit 9, lot 954 recall).",
      "recall_initiation_date": "20151221",
      "center_classification_date": "20160308",
      "termination_date": "20180530",
      "report_date": "20160316",
      "code_info": "Software Version 9.4 and below; VITROS 250 Serial Numbers: 25010043 - 25055306; VITROS 250 Refurbished Serial Numbers: 25010582, 25010617, 25010638, 25010782, 25011056, 25011074, 25011176, 25011192, 25011196, 25011246, 25011317, 25011383, 25011408, 25011435, 25011448, 25011517, 25011647, 25011648, 25011777, 25011957, 25012025, 25012058, 25012062, 25012073, 25012078, 25012140, 25012259, 25012267, 25012323, 25012344, 25012487, 25012505, 25012521, 25012574, 25012629, 25012654, 25012672, 25012765, 25012782, 25012838, 25012931, 25012938, 25013036, 25013100, 25013139, 25013216, 25013245, 25013313, 25013389, 25013391, 25013448, 25013473, 25013497, 25013583, 25013597, 25013719, 25013757, 25013787, 25013790, 25013870, 25013879, 25013971, 25013982, 25014033, 25014196, 25014201, 25014209, 25014272, 25014278, 25014286, 25014318, 25014404, 25014409, 25014412, 25014448, 25014449, 25014459, 25014562, 25014609, 25014610, 25014655, 25014657, 25014663, 25014728, 25014741, 25014775, 25014788, 25014832, 25014849, 25014927, 25015076, 25015132, 25015133, 25015169, 25015223, 25015242, 25015277, 25015389, 25015430, 25015445, 25015463, 25015678, 25015770, 25015847, 25015914, 25015915, 25015976, 25015990, 25015998, 25016011, 25016177, 25016218, 25016429, 25016463, 25016517, 25016528, 25016552, 27000025, 27000054, 27000132, 27000156, 27000216, 27000317, 27000488, 27000870, 27000936, 27001058, 27001124, 27001471, 27001644, 27001651, 27001763, 27002256, 27002351, 27002714, R2500159, R2500179, R2500217, 25010122, 25010187, 25010220, 25010320, 25010332, 25010360, 25010362, 25010380, 25010394, 25010429, 25010450, 25010465, 25010522, 25010584, 25010648, 25010677, 25010690, 25010691, 25010694, 25010697, 25010709, 25010713, 25010719, 25010725, 25010735, 25010749, 25010751, 25010753, 25010770, 25010772, 25010774, 25010775, 25010802, 25010809, 25010817, 25010841, 25010845, 25010848, 25010850, 25010859, 25010864, 25010865, 25010867, 25010872, 25010885, 25010889, 25010908, 25010919, 25010923, 25010949, 25010954, 25010955, 25010962, 25010965, 25010970, 25010996, 25011018, 25011030, 25011031, 25011043, 25011061, 25011072, 25011078, 25011083, 25011107, 25011109, 25011126, 25011131, 25011133, 25011161, 25011179, 25011187, 25011205, 25011208, 25011230, 25011231, 25011243, 25011247, 25011255, 25011256, 25011274, 25011278, 25011283, 25011324, 25011341, 25011345, 25011356, 25011361, 25011362, 25011372, 25011384, 25011385, 25011388, 25011390, 25011392, 25011393, 25011407, 25011424, 25011436, 25011441, 25011446, 25011456, 25011462, 25011466, 25011473, 25011476, 25011479, 25011482, 25011500, 25011501, 25011508, 25011513, 25011521, 25011525, 25011533, 25011535, 25011541, 25011553, 25011556, 25011565, 25011569, 25011599, 25011610, 25011619, 25011624, 25011627, 25011631, 25011649, 25011650, 25011673, 25011678, 25011679, 25011684, 25011698, 25011701, 25011709, 25011710, 25011712, 25011719, 25011729, 25011732, 25011743, 25011749, 25011751, 25011759, 25011780, 25011783, 25011785, 25011795, 25011797, 25011816, 25011817, 25011826, 25011834, 25011836, 25011845, 25011852, 25011854, 25011868, 25011872, 25011878, 25011880, 25011881, 25011887, 25011891, 25011900, 25011904, 25011922, 25011937, 25011940, 25011949, 25011956, 25011960, 25011964, 25011965, 25011966, 25011968, 25011983, 25011985, 25011987, 25011988, 25012001, 25012002, 25012004, 25012015, 25012016,   25012017, 25012024, 25012029, 25012042, 25012043, 25012044, 25012052, 25012053, 25012056, 25012065, 25012066, 25012074, 25012087, 25012090, 25012102, 25012105, 25012111, 25012115, 25012118, 25012127, 25012134, 25012155, 25012169, 25012192, 25012193, 25012196, 25012221, 25012227, 25012231, 25012242, 25012262, 25012264, 25012270, 25012274, 25012283, 25012298, 25012301, 25012302, 25012310, 25012312, 25012313, 25012325, 25012336, 25012351, 25012357, 25012360, 25012371, 25012379, 25012380, 25012391, 25012404, 25012411, 25012418, 25012421, 25012422, 25012429, 25012431, 25012436, 25012444, 25012445, 25012464, 25012467, 25012478, 25012479, 25012480, 25012481, 25012482, 25012483, 25012485, 25012489, 25012494, 25012504, 25012506, 25012511, 25012512, 25012514, 25012525, 25012545, 25012547, 25012552, 25012579, 25012582, 25012597, 25012614, 25012624, 25012626, 25012628, 25012641, 25012648, 25012651, 25012652, 25012653, 25012655, 25012656, 25012657, 25012658, 25012660, 25012661, 25012663, 25012680, 25012707, 25012710, 25012718, 25012723, 25012726, 25012728, 25012733, 25012735, 25012736, 25012742, 25012751, 25012775, 25012779, 25012784, 25012786, 25012791, 25012793, 25012803, 25012806, 25012808, 25012821, 25012836, 25012843, 25012861, 25012878, 25012879, 25012883, 25012886, 25012888, 25012897, 25012903, 25012906, 25012914, 25012921, 25012924, 25012927, 25012930, 25012937, 25012939, 25012941, 25012943, 25012950, 25012951, 25012952, 25012956, 25012966, 25012971, 25012974, 25012976, 25012985, 25012988, 25012995, 25013001, 25013002, 25013006, 25013009, 25013010, 25013015, 25013019, 25013020, 25013021, 25013025, 25013033, 25013037, 25013042, 25013045, 25013054, 25013063, 25013068, 25013073, 25013074, 25013075, 25013091, 25013110, 25013114, 25013119, 25013124, 25013151, 25013155, 25013157, 25013174, 25013185, 25013205, 25013207, 25013210, 25013219, 25013222, 25013230, 25013236, 25013239, 25013241, 25013244, 25013248, 25013258, 25013265, 25013272, 25013273, 25013278, 25013279, 25013282, 25013284, 25013297, 25013303, 25013312, 25013314, 25013315, 25013316, 25013318, 25013331, 25013332, 25013334, 25013335, 25013336, 25013354, 25013362, 25013364, 25013374, 25013376, 25013383, 25013386, 25013388, 25013393, 25013398, 25013413, 25013416, 25013424, 25013440, 25013441, 25013443, 25013444, 25013456, 25013459, 25013472, 25013478, 25013483, 25013484, 25013488, 25013491, 25013492, 25013496, 25013505, 25013508, 25013511, 25013522, 25013531, 25013539, 25013544, 25013545, 25013549, 25013558, 25013559, 25013561, 25013562, 25013563, 25013564, 25013568, 25013573, 25013580, 25013581, 25013582, 25013584, 25013589, 25013592, 25013599, 25013601, 25013603, 25013612, 25013615, 25013617, 25013622, 25013623, 25013631, 25013632, 25013637, 25013659, 25013662, 25013665, 25013666, 25013668, 25013669, 25013670, 25013673, 25013675, 25013679, 25013689, 25013693, 25013694, 25013697, 25013707, 25013713, 25013715, 25013717, 25013718, 25013732, 25013733, 25013736, 25013737, 25013739, 25013747, 25013755, 25013766, 25013772, 25013773, 25013775, 25013776, 25013779, 25013793, 25013812, 25013814, 25013820, 25013824, 25013825, 25013826, 25013834, 25013840,   25013844, 25013847, 25013859, 25013862, 25013867, 25013869, 25013873, 25013876, 25013878, 25013881, 25013882, 25013887, 25013888, 25013890, 25013894, 25013900, 25013904, 25013907, 25013908, 25013909, 25013915, 25013916, 25013917, 25013932, 25013935, 25013938, 25013940, 25013945, 25013954, 25013957, 25013958, 25013959, 25013967, 25013968, 25013969,   25013972, 25013979, 25013980, 25013987, 25013988, 25013989, 25013994, 25014011, 25014012, 25014016, 25014020, 25014024, 25014031, 25014037, 25014047, 25014050, 25014051, 25014055, 25014057, 25014060, 25014065, 25014067, 25014072, 25014074, 25014087, 25014093, 25014097, 25014100, 25014109, 25014119, 25014121, 25014126, 25014127, 25014129, 25014130,   25014131, 25014137, 25014149, 25014155, 25014169, 25014172, 25014173,  25014174, 25014182, 25014191, 25014193, 25014194, 25014199, 25014211, 25014222, 25014223, 25014226, 25014236, 25014239, 25014244, 25014245, 25014250, 25014252, 25014260, 25014263, 25014265, 25014274, 25014276, 25014277, 25014279, 25014281, 25014288, 25014309, 25014310, 25014311,   25014319, 25014322, 25014325, 25014330, 25014343, 25014357, 25014366, 25014376, 25014378, 25014380, 25014384, 25014386, 25014390, 25014391, 25014392, 25014425, 25014426, 25014432, 25014435, 25014441, 25014452, 25014467, 25014468, 25014469, 25014477, 25014479, 25014489, 25014503, 25014507, 25014517, 25014518, 25014529, 25014539, 25014541, 25014551,   25014566, 25014567, 25014573, 25014578, 25014580, 25014582, 25014603, 25014604, 25014613, 25014617, 25014620, 25014627, 25014628, 25014630, 25014631, 25014638, 25014650, 25014654, 25014674, 25014677, 25014696, 25014697, 25014707, 25014714, 25014715, 25014716, 25014717, 25014719, 25014725, 25014730, 25014737, 25014738, 25014739, 25014747, 25014763,  25014765, 25014767, 25014769, 25014771, 25014783, 25014787, 25014789, 25014796, 25014805, 25014807, 25014821, 25014823, 25014824, 25014826, 25014827, 25014831, 25014834, 25014838, 25014840, 25014841, 25014843, 25014844, 25014851, 25014855, 25014856, 25014857, 25014858, 25014865, 25014866, 25014877, 25014878, 25014886, 25014887, 25014888, 25014889,   25014893, 25014896, 25014899, 25014902, 25014905, 25014908, 25014923, 25014924, 25014935, 25014938, 25014939, 25014940, 25014942, 25014947, 25014951, 25014952, 25014958, 25014970, 25014973, 25014974, 25014977, 25014979, 25014984, 25014986, 25014992, 25014994, 25014995, 25014998, 25015004, 25015013, 25015014, 25015019, 25015029, 25015031, 25015039,   25015042, 25015044, 25015048, 25015053, 25015055, 25015063, 25015068, 25015075, 25015086, 25015090, 25015093, 25015099, 25015105, 25015108, 25015112, 25015120, 25015126, 25015137, 25015139, 25015140, 25015141, 25015147, 25015160, 25015161, 25015165, 25015171, 25015192, 25015205, 25015215, 25015216, 25015218, 25015225, 25015231, 25015235, 25015236,   25015239, 25015240, 25015249, 25015256, 25015257, 25015259, 25015261, 25015267, 25015274, 25015284, 25015291, 25015293, 25015294, 25015296, 25015299, 25015307, 25015310, 25015311, 25015318, 25015319, 25015321, 25015323, 25015326, 25015338, 25015339, 25015342, 25015350, 25015351, 25015353, 25015354, 25015355, 25015358, 25015359, 25015360, 25015367, 25015372, 25015373, 25015378, 25015381, 25015382, 25015386, 25015387, 25015393, 25015404, 25015405, 25015408, 25015413, 25015415, 25015416, 25015424, 25015425, 25015433, 25015435, 25015441, 25015452, 25015467, 25015477, 25015483, 25015484, 25015487, 25015503, 25015513, 25015518, 25015521, 25015529, 25015559, 25015566, 25015577, 25015578, 25015585,   25015589, 25015590, 25015597, 25015600, 25015603, 25015610, 25015611, 25015612, 25015614, 25015625, 25015633, 25015634, 25015639, 25015641, 25015644, 25015645, 25015650, 25015654, 25015658, 25015661, 25015670, 25015671, 25015674, 25015676, 25015682, 25015685, 25015690, 25015699, 25015702, 25015704, 25015705, 25015710, 25015712, 25015713, 25015717,   25015723, 25015734, 25015736, 25015739, 25015741, 25015742, 25015757, 25015763, 25015764, 25015765, 25015782, 25015783, 25015787, 25015788, 25015789, 25015803, 25015807, 25015808, 25015811, 25015812, 25015814, 25015817, 25015824, 25015825, 25015837, 25015842, 25015846, 25015851, 25015856, 25015859, 25015860, 25015863, 25015871, 25015874, 25015875,   25015878, 25015879, 25015882, 25015889, 25015894, 25015898, 25015904, 25015912, 25015913, 25015923, 25015924, 25015935, 25015938, 25015946, 25015958, 25015963, 25015965, 25015969, 25015970, 25015973, 25015981, 25015983, 25015985, 25016000, 25016009, 25016012, 25016019, 25016021, 25016030, 25016037, 25016039, 25016041, 25016045, 25016048, 25016049, 25016051, 25016053, 25016054, 25016058, 25016063, 25016069, 25016073, 25016086, 25016087, 25016088, 25016095, 25016102, 25016104, 25016108, 25016113, 25016115, 25016116, 25016118, 25016120, 25016126, 25016128, 25016142, 25016149, 25016151, 25016159, 25016164, 25016172, 25016176, 25016181, 25016215, 25016231, 25016234, 25016240, 25016241, 25016242,   25016255, 25016268, 25016273, 25016276, 25016290, 25016298, 25016299, 25016301, 25016305, 25016309, 25016316, 25016324, 25016325, 25016333,  25016337, 25016357, 25016360, 25016361, 25016368, 25016384, 25016391, 25016392, 25016404, 25016407, 25016411, 25016414, 25016418, 25016426, 25016428, 25016431, 25016433, 25016437, 25016441, 25016443, 25016446,   25016452, 25016453, 25016458, 25016461, 25016467, 25016474, 25016479, 25016480, 25016486, 25016490, 25016492, 25016494, 25016495, 25016497, 25016511, 25016514, 25016524, 25016549, 25016551, 25016556, 25016557, 25016558, 25016563, 25016566, 25016573, 25016580, 25016582, 25016586, 25016589, 25016594, 25016595, 25016596, 25016602, 25016604, 25016613,   25016619, 25016620, 25016637, 25016638, 25016644, 25016647, 25016649, 25016650, 25016670, 25016804, 25016818, 25016837, 25016838, 25016852, 25016899, 25016912, 25016916, 25016917, 25016918, 25016919, 25016932, 25016940, 25016943, 25016944, 25016946, 25016947, 25016950, 25016951, 25016952, 25016953, 25016957, 25016958, 25104774, 27000004, 27000006,   27000007, 27000008, 27000009, 27000010, 27000016, 27000018, 27000022, 27000041, 27000051, 27000052, 27000058, 27000061, 27000063, 27000070, 27000072, 27000075, 27000080, 27000089, 27000103, 27000106, 27000110, 27000112, 27000115, 27000126, 27000133, 27000134, 27000135, 27000136, 27000146, 27000147, 27000148, 27000164, 27000165, 27000175, 27000181,   27000183, 27000192, 27000203, 27000205, 27000206, 27000208, 27000209, 27000217, 27000223, 27000224, 27000227, 27000233, 27000244, 27000273, 27000281, 27000287, 27000289, 27000290, 27000300, 27000303, 27000304, 27000309, 27000313, 27000316, 27000324, 27000326, 27000328, 27000333, 27000350, 27000351, 27000355, 27000357, 27000362, 27000369, 27000372,   27000373, 27000374, 27000382, 27000383, 27000384, 27000389, 27000409, 27000415, 27000418, 27000421, 27000424, 27000427, 27000436, 27000448, 27000450, 27000453, 27000456, 27000457, 27000467, 27000470, 27000471, 27000477, 27000479, 27000483, 27000484, 27000489, 27000492, 27000507, 27000515, 27000516, 27000529, 27000538, 27000540, 27000546, 27000568,   27000581, 27000584, 27000598, 27000606, 27000607, 27000624, 27000627, 27000652, 27000655, 27000658, 27000660, 27000662, 27000663, 27000677, 27000678, 27000680, 27000683, 27000686, 27000696, 27000709, 27000713, 27000714, 27000716, 27000717, 27000718, 27000719, 27000740, 27000745, 27000758, 27000769, 27000777, 27000779, 27000784, 27000787, 27000790,   27000791, 27000796, 27000797, 27000806, 27000813, 27000814, 27000824, 27000835, 27000847, 27000854, 27000861, 27000862, 27000869, 27000872, 27000873, 27000874, 27000875, 27000877, 27000880, 27000881, 27000882, 27000894, 27000903, 27000908, 27000912, 27000924, 27000926, 27000927, 27000928, 27000930, 27000947, 27000948, 27000949, 27000958, 27000961,  27000963, 27000964, 27000984, 27000988, 27000994, 27001005, 27001006, 27001007, 27001021, 27001032, 27001061, 27001066, 27001087, 27001089, 27001091, 27001093, 27001094, 27001103, 27001104, 27001123, 27001139, 27001144, 27001145, 27001166, 27001167, 27001170, 27001177, 27001183, 27001185, 27001187, 27001193, 27001202, 27001205, 27001212, 27001213,  27001221, 27001223, 27001229, 27001234, 27001236, 27001240, 27001241, 27001279, 27001283, 27001289, 27001304, 27001319, 27001326, 27001334, 27001337, 27001353, 27001360, 27001361, 27001364, 27001366, 27001381, 27001400, 27001402, 27001418, 27001426, 27001429, 27001434, 27001448, 27001473, 27001497, 27001520, 27001524, 27001525, 27001526, 27001531,   27001533, 27001535, 27001552, 27001556, 27001559, 27001565, 27001599, 27001612, 27001639, 27001645, 27001646, 27001647, 27001660, 27001662, 27001667, 27001670, 27001692, 27001693, 27001694, 27001696, 27001698, 27001729, 27001741, 27001766, 27001772, 27001785, 27001796, 27001802, 27001806, 27001871, 27001901, 27001905, 27001912, 27001934, 27001952,   27001953, 27001968, 27001975, 27002013, 27002018, 27002028, 27002044, 27002134, 27002136, 27002151, 27002183, 27002245, 27002304, 27002350, 27002352, 27002521, 27002526, 27002557, 27002652, 27002663, 25010717A, R2500004, R2500011, R2500028, R2500033, R2500037, R2500041, R2500044, R2500050, R2500063, R2500081, R2500087, R2500125, R2500140, R2500146, R2500147, R2500149, R2500155, R2500158, R2500173, R2500183, R2500188, R2500189, R2500191, R2500232"
    }
  ]
}