  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Tustin",
      "address_1": "2441 Michelle Dr",
      "reason_for_recall": "Due to a problem with the control software of the x-ray high-voltage generator, if scanning is performed with a tube current less than 30mA, the control software may detect an error and perform error processing even though the x-ray output is normal.",
      "address_2": "P.O. Box 2068",
      "product_quantity": "1 unit",
      "code_info": "Serial No: 1CB1262006",
      "center_classification_date": "20130408",
      "distribution_pattern": "Distributed in the state of MD.",
      "state": "CA",
      "product_description": "Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning.",
      "report_date": "20130417",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Toshiba American Medical Systems Inc",
      "recall_number": "Z-1068-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64673",
      "termination_date": "20130408",
      "more_code_info": "",
      "recall_initiation_date": "20121114",
      "postal_code": "92780-7047",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}