  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "El Segundo",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89880",
      "recalling_firm": "Karl Storz Endoscopy",
      "address_1": "2151 E Grand Ave",
      "address_2": "N/A",
      "postal_code": "90245-5017",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA,WI, WV, and WY.  O.U.S.: Mexico, Lebanon, and South Korea",
      "recall_number": "Z-1067-2022",
      "product_description": "11278A2 11278AK2 Flexible Ureteroscope Z20615US-BA (08-2018);  11278AC1 N/A Flexible Ureteroscope Z17859US-A (04/2016);  11278AC2 11278ACK2 Flexible Ureteroscope Z20615US-BA (08-2018);  11278ACU1 11278ACUK1 Flexible Ureteroscope Z20615US-BA (08-2018);  11278AU1 11278AUK1 Flexible Ureteroscope Z20615US-BA (08-2018);",
      "product_quantity": "All Serial Numbers manufactured/distributed since January 2018.",
      "reason_for_recall": "Failure to achieve the expected six-log reduction in microorganisms following the disinfection process.",
      "recall_initiation_date": "20220401",
      "center_classification_date": "20220512",
      "report_date": "20220518",
      "code_info": "11278A2 11278AK2 Flexible Ureteroscope Z20615US-BA (08-2018) UDI Code: 4048551402573; 11278AC1 N/A Flexible Ureteroscope Z17859US-A (04/2016) UDI Code: 4048551226353; 11278AC2 11278ACK2 Flexible Ureteroscope Z20615US-BA (08-2018) UDI Code: 4048551402504; 11278ACU1 11278ACUK1 Flexible Ureteroscope Z20615US-BA (08-2018) UDI Code: 4048551226377; 11278AU1 11278AUK1 Flexible Ureteroscope Z20615US-BA (08-2018) UDI Code: 4048551226438;  All Serial Numbers manufactured/distributed since January 2018."
    }
  ]
}