  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Waltham",
      "address_1": "500 Totten Pond Rd",
      "reason_for_recall": "Certain lots are missing a protective sheath component over the silicone tip, which may result in the silicone tip breaking or detaching at the metal tube component. If the tip becomes detached during a procedure while in the eye, the fragment will need to be retrieved to mitigate the possibility of injury.",
      "address_2": "10 City Point",
      "product_quantity": "3430 units",
      "code_info": "Lots 60047S4, 60060S9, 600623S",
      "center_classification_date": "20190327",
      "distribution_pattern": "Worldwide distribution: US (nationwide) to states of:  AR, AZ, CA, CO, FL, GA, LA, MD, MO, NC, NE, NJ, NY, OH, OK, TX, UT, VA, WA; and countries of: Mexico and Peru.",
      "state": "MA",
      "product_description": "BVI Visitec Soft Tip Cannula, 0.8MM .50 x 33mm (25G x 1.3 in).    The BVI Soft Tip Cannula is intended to irrigate and/or aspirate ophthalmic solutions and liquids in or out of the eye, and remove materials where appropriate (e.g. retinal, subretinal procedures).",
      "report_date": "20190403",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Beaver Visitec",
      "recall_number": "Z-1066-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "81757",
      "termination_date": "20200501",
      "recall_initiation_date": "20181130",
      "postal_code": "02451-1916",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}