      "status": "Ongoing",
      "city": "Oxford",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73459",
      "recalling_firm": "Villa Radiology Systems LLC",
      "address_1": "91 Willenbrock Rd Ste B1",
      "address_2": "N/A",
      "postal_code": "06478-1036",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "AL  AZ  CA  CO  CT  DC  FL  IA  IL  IN  LA  MA  MD  MI  MN  MO  MS  NJ  NY  OH  PA  Puerto Rico  TN  TX  UT  WI",
      "recall_number": "Z-1064-2017",
      "product_description": "Apollo: Models: 9784000131, 9784000231, 9784020131, 9784020231, 9784100231, 9784120131, 9784120231, 9784130131    Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.",
      "product_quantity": "66 distributed in US",
      "reason_for_recall": "The analysis showed that the incident was caused by a chain of four faults:  1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position.  2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area.  3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation).  4.) Delayed activation of the emergency stop.",
      "recall_initiation_date": "20160219",
      "center_classification_date": "20170120",
      "report_date": "20170201",
      "code_info": "06040078  07030172  06090116  06110138  07010149  07020158  07020162  07020163  07020165  07030168  07030171  07040206  09090605  10100752  06060086  06100127  09010566  08020473  11100873  15041446"
    }
  ]
}